Comparing real-bone measurement microphone and skull simulator measurements for soft headband bone conduction devices in adults

Introduction The use of bone conduction devices has increased in recent years, offering effective solutions for patients with conductive or mixed hearing loss and limited usable hearing unilaterally. This growth necessitates more accurate verification tools to ensure optimal patient outcomes. Methods The real-bone measurement microphone is an innovative technology that enables in-situ verification of various bone conduction device types and is under development by Audioscan. This study presents measurements made using this technology on 10 adults wearing two commercially available soft headband bone conduction devices. Comparable measurements were conducted with the Verifit Skull Simulator, a commercially available technology for verifying abutment-based bone conduction devices. Force levels recorded with both technologies are compared. Results Results include real-head-to-dial difference and dB force level measurements across various scenarios. Average real-head-to-dial difference values measured using two bone conduction devices differ at most frequencies, while following the same overall trend as frequency changes. For the dB force level results obtained from the real-bone measurement microphone and Verifit Skull Simulator, the Pearson correlation between these two devices is consistently greater than 0.87, and the root mean square error varies between 6 to 12 dB across different test scenarios. The potential sources of differences between the two devices are analyzed and discussed in this paper. Conclusion Findings provide insights into the performance characteristics of both technologies, demonstrating that the real-bone measurement microphone is a promising tool for expanding clinical verification capabilities across a wide range of bone conduction devices.

Authors

Institutions

Publication Details

Journal
Frontiers in Audiology and Otology
Published
2026-04-22
DOI
https://doi.org/10.3389/fauot.2026.1768441
Primary Topic
Ear Surgery and Otitis Media
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparing real-bone measurement microphone and skull simulator measurements for soft headband bone conduction devices in adults

John Pumford, Arash Ebrahimian, Neritan Alizoti, Chris McInerney et al.
Frontiers in Audiology and Otology
Ear Surgery and Otitis Media
article

Comparing real-bone measurement microphone and skull simulator measurements for soft headband bone conduction devices in adults

John Pumford, Arash Ebrahimian, Neritan Alizoti, Chris McInerney, Jonathan Pietrobon, Matthew Lucas
article en

Abstract

Introduction The use of bone conduction devices has increased in recent years, offering effective solutions for patients with conductive or mixed hearing loss and limited usable hearing unilaterally. This growth necessitates more accurate verification tools to ensure optimal patient outcomes. Methods The real-bone measurement microphone is an innovative technology that enables in-situ verification of various bone conduction device types and is under development by Audioscan. This study presents measurements made using this technology on 10 adults wearing two commercially available soft headband bone conduction devices. Comparable measurements were conducted with the Verifit Skull Simulator, a commercially available technology for verifying abutment-based bone conduction devices. Force levels recorded with both technologies are compared. Results Results include real-head-to-dial difference and dB force level measurements across various scenarios. Average real-head-to-dial difference values measured using two bone conduction devices differ at most frequencies, while following the same overall trend as frequency changes. For the dB force level results obtained from the real-bone measurement microphone and Verifit Skull Simulator, the Pearson correlation between these two devices is consistently greater than 0.87, and the root mean square error varies between 6 to 12 dB across different test scenarios. The potential sources of differences between the two devices are analyzed and discussed in this paper. Conclusion Findings provide insights into the performance characteristics of both technologies, demonstrating that the real-bone measurement microphone is a promising tool for expanding clinical verification capabilities across a wide range of bone conduction devices.

Frontiers in Audiology and OtologyVol. 4
Western University (CA), Protected Trust Services (United Kingdom) (GB)
Openalex Percentile: Top 14%
Ear Surgery and Otitis Media
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.