Applications of a 32-Channel Ultra-flexible Phased-Array Prototype Coil Design Optimized for Small Joint 3-Tesla MRI

Objective 
MRI of the hand and wrist region, containing small structures such as ligaments, peripheral nerves, and thin cartilage, is challenging, requiring both high spatial resolution and coverage, along with adequate signal-to-noise ratio (SNR) for image quality. This work aimed to evaluate whether a high-density, prototype flexible coil array could better visualize small anatomic structures without compromising SNR or field-of-view (FOV), compared to conventional hand and wrist coils. 

Approach 
A flexible, 32-channel phased-array prototype coil was developed for 3T hand and wrist MRI using novel, small dual-loop elements. A silicon oil phantom was used to analyze SNR and noise amplification arising from acceleration factors (R) of 2-5. Confirmation of the prototype's phantom performance was done in wrist MRI comparisons in two healthy volunteers and against two 16-channel commercially available coils for superior-inferior coverage and at different acceleration factors (R=1.5-4). Additional imaging in 6 patients was performed in the hand, wrist, elbow, brachial plexus, and foot for anecdotal, qualitative assessment of the coil's performance. 

Main Results 
Phantom testing demonstrated ~26% higher SNR and 19.2%-192.4% lower g-factors in the prototype compared to a 16-channel conventional coil. In comparisons against other coils, the prototype coil demonstrated larger superior-inferior coverage of 23.9 cm compared to 20.5 cm and 20.6 cm in conventional coils. Image quality was maintained at higher acceleration factors. Patient imaging successfully demonstrated visualization of small structures. 

Significance 
The 32-channel prototype coil's extended FOV, improved SNR, and flexible design enabled accelerated scans with high spatial resolution and broader anatomical coverage, offering technical improvements over conventional coil designs.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-04-22
DOI
https://doi.org/10.1088/1361-6560/ae639d
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Applications of a 32-Channel Ultra-flexible Phased-Array Prototype Coil Design Optimized for Small Joint 3-Tesla MRI

Yun-Jeong Stickle, Ek T. Tan, Fraser Robb, Maggie Fung et al.
Physics in Medicine and Biology
Advanced MRI Techniques and Applications
article

Applications of a 32-Channel Ultra-flexible Phased-Array Prototype Coil Design Optimized for Small Joint 3-Tesla MRI

Yun-Jeong Stickle, Ek T. Tan, Fraser Robb, Maggie Fung, Yan Wen, Darryl B. Sneag, Allison Sydney Lowe, Cassiano Santiago
article en

Abstract

Objective 
MRI of the hand and wrist region, containing small structures such as ligaments, peripheral nerves, and thin cartilage, is challenging, requiring both high spatial resolution and coverage, along with adequate signal-to-noise ratio (SNR) for image quality. This work aimed to evaluate whether a high-density, prototype flexible coil array could better visualize small anatomic structures without compromising SNR or field-of-view (FOV), compared to conventional hand and wrist coils. 

Approach 
A flexible, 32-channel phased-array prototype coil was developed for 3T hand and wrist MRI using novel, small dual-loop elements. A silicon oil phantom was used to analyze SNR and noise amplification arising from acceleration factors (R) of 2-5. Confirmation of the prototype's phantom performance was done in wrist MRI comparisons in two healthy volunteers and against two 16-channel commercially available coils for superior-inferior coverage and at different acceleration factors (R=1.5-4). Additional imaging in 6 patients was performed in the hand, wrist, elbow, brachial plexus, and foot for anecdotal, qualitative assessment of the coil's performance. 

Main Results 
Phantom testing demonstrated ~26% higher SNR and 19.2%-192.4% lower g-factors in the prototype compared to a 16-channel conventional coil. In comparisons against other coils, the prototype coil demonstrated larger superior-inferior coverage of 23.9 cm compared to 20.5 cm and 20.6 cm in conventional coils. Image quality was maintained at higher acceleration factors. Patient imaging successfully demonstrated visualization of small structures. 

Significance 
The 32-channel prototype coil's extended FOV, improved SNR, and flexible design enabled accelerated scans with high spatial resolution and broader anatomical coverage, offering technical improvements over conventional coil designs.

Physics in Medicine and Biology
Hospital for Special Surgery (US), General Electric (Spain) (ES), Applied Sciences Laboratory (United States) (US)
GE Healthcare
Openalex Percentile: Top 13%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.