Toward XRF-like chemical sensitivity in the laboratory: first hyperspectral x-ray imaging demonstration using the novel CITIUS detector

Fe source, a sharp energy resolution in the 0.5-0.8 keV range (full width at half maximum) was found, enabling fine energy binning suitable for advanced quantitative material identification. For the first demonstration, planar and tomographic datasets of two multi-material test samples were analyzed using a newly adapted version of the Minimum-Residual Basis Material Decomposition (MR-BMD) algorithm, optimized for hyperspectral detectors providing tens of energy bins with narrow bandwidths. Results show that the laboratory-based hyperspectral approach combined with MR-BMD enables element-sensitive imaging and, remarkably, separates materials with very similar attenuation, such as water and polypropylene. These results demonstrate accurate material identification and quantification, promisingly approaching XRF-like chemical sensitivity in the laboratory.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-04-22
DOI
https://doi.org/10.1038/s41598-026-48904-6
Primary Topic
Advanced X-ray Imaging Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Toward XRF-like chemical sensitivity in the laboratory: first hyperspectral x-ray imaging demonstration using the novel CITIUS detector

Y. Honjo, T. Hatsui, H. Nishino, V. Di Trapani et al.
Scientific Reports
Advanced X-ray Imaging Techniques
article

Toward XRF-like chemical sensitivity in the laboratory: first hyperspectral x-ray imaging demonstration using the novel CITIUS detector

Y. Honjo, T. Hatsui, H. Nishino, V. Di Trapani, F. Orsini, F. Arfelli, R. H. Menk, Y. Joti, P. Thibault, K. Ozaki
article en

Abstract

Fe source, a sharp energy resolution in the 0.5-0.8 keV range (full width at half maximum) was found, enabling fine energy binning suitable for advanced quantitative material identification. For the first demonstration, planar and tomographic datasets of two multi-material test samples were analyzed using a newly adapted version of the Minimum-Residual Basis Material Decomposition (MR-BMD) algorithm, optimized for hyperspectral detectors providing tens of energy bins with narrow bandwidths. Results show that the laboratory-based hyperspectral approach combined with MR-BMD enables element-sensitive imaging and, remarkably, separates materials with very similar attenuation, such as water and polypropylene. These results demonstrate accurate material identification and quantification, promisingly approaching XRF-like chemical sensitivity in the laboratory.

Scientific Reports
University of Trieste (IT), Elettra-Sincrotrone Trieste S.C.p.A. (IT), SPring-8 (JP), Istituto Nazionale di Fisica Nucleare, Sezione di Trieste (IT), Mid Sweden University (SE)
European Research Council
Affordable and clean energy
Openalex Percentile: Top 8%
Advanced X-ray Imaging Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.