A nonmagnetic micropositioner for displacement measurement validation in the MRI environment
OBJECTIVE: The purpose of this study was to implement a nonmagnetic micropositioner in the MRI environment to validate displacement estimates of magnetic resonance acoustic radiation force imaging (MR-ARFI). METHODS: The micropositioner consisted of a stage driven by a piezoelectric stepper motor in closed-loop operation with an optical quadrature encoder. A 100-gram agar gel phantom was prepared, and three MR-ARFI pulse sequences were used to generate displacement maps. MR-ARFI measured displacements were compared to ground truth data from the optical encoder. RESULTS: but exhibited scaling biases that may have been due to the experimental setup. CONCLUSION: The proposed instrument can potentially improve the accuracy and precision of MR-ARFI-based applications, including focused ultrasound dosimetry and mechanical biomarker imaging.
Authors
- William R. Willoughby (ORCID: https://orcid.org/0000-0002-0773-2453)
- Mark Bolding
Institutions
- University of Alabama at Birmingham (US)
Publication Details
- Journal
- Magnetic Resonance Materials in Physics Biology and Medicine
- Published
- 2026-04-21
- DOI
- https://doi.org/10.1007/s10334-026-01341-y
- Primary Topic
- Ultrasound Imaging and Elastography
- Type
- article
- Field-Weighted Citation Impact
- 0.00