High-resolution bone imaging and spectral soft-tissue contrast in post-mortem photon-counting CT

• Photon-counting CT shows lower image noise than energy-integrating CT post mortem • High-resolution photon-counting CT improves bone and lung detail • Spectral imaging enables organ-specific soft-tissue contrast optimization • Targeted reconstructions add value without requiring additional scan acquisitions Photon-counting detector (PCD) computed tomography (CT) offers reduced image noise, high spatial resolution, and intrinsic spectral imaging capabilities, making it a promising technology for post-mortem CT. However, the relative contribution of PCD-CT to post-mortem imaging still needs to be characterized. This short communication provides a stepwise evaluation of key determinants of image quality in PMCT using a controlled post-mortem animal model. Whole-body CT was performed on a single post-mortem animal specimen using a PCD CT system and a conventional energy-integrating detector (EID) CT system under matched acquisition and reconstruction conditions. Image quality was evaluated sequentially by comparing background noise characteristics between detector types, assessing high-resolution photon-counting CT protocols for osseous structures, and analyzing soft-tissue image quality using SNR and CNR under both polychromatic and virtual mono-energetic reconstruction conditions. PCD CT demonstrated significantly lower background noise compared with EID CT ( p < 0.01). High-resolution PCD CT markedly improved visualization of fine osseous and pulmonary structures, while no consistent qualitative benefit for soft tissue was visually observed. Under matched polychromatic conditions, soft-tissue contrast improvements were modest and tissue-dependent. The most pronounced soft-tissue advantages were observed with spectral imaging, where virtual mono-energetic reconstructions showed a strong energy-dependent behavior. CNR improvements were organ-specific, with clear benefits for renal parenchyma but not for liver tissue. These findings indicate that the principal added value of PCD CT in post-mortem imaging is task-dependent. Detector-related noise reduction provides a robust baseline benefit, high-resolution imaging primarily enhances osseous assessment, and spectral reconstructions offer the greatest potential for soft-tissue optimization.

Authors

Institutions

Publication Details

Journal
Forensic Imaging
Published
2026-04-21
DOI
https://doi.org/10.1016/j.fri.2026.200818
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High-resolution bone imaging and spectral soft-tissue contrast in post-mortem photon-counting CT

Sutter Reto, Gascho Dominic, Michael Josef Thali, Beckmann M. Katrin et al.
Forensic Imaging
Advanced X-ray and CT Imaging
article

High-resolution bone imaging and spectral soft-tissue contrast in post-mortem photon-counting CT

Sutter Reto, Gascho Dominic, Michael Josef Thali, Beckmann M. Katrin, Zoelch Niklaus, Eva Deininger-Czermak
article en

Abstract

• Photon-counting CT shows lower image noise than energy-integrating CT post mortem • High-resolution photon-counting CT improves bone and lung detail • Spectral imaging enables organ-specific soft-tissue contrast optimization • Targeted reconstructions add value without requiring additional scan acquisitions Photon-counting detector (PCD) computed tomography (CT) offers reduced image noise, high spatial resolution, and intrinsic spectral imaging capabilities, making it a promising technology for post-mortem CT. However, the relative contribution of PCD-CT to post-mortem imaging still needs to be characterized. This short communication provides a stepwise evaluation of key determinants of image quality in PMCT using a controlled post-mortem animal model. Whole-body CT was performed on a single post-mortem animal specimen using a PCD CT system and a conventional energy-integrating detector (EID) CT system under matched acquisition and reconstruction conditions. Image quality was evaluated sequentially by comparing background noise characteristics between detector types, assessing high-resolution photon-counting CT protocols for osseous structures, and analyzing soft-tissue image quality using SNR and CNR under both polychromatic and virtual mono-energetic reconstruction conditions. PCD CT demonstrated significantly lower background noise compared with EID CT ( p < 0.01). High-resolution PCD CT markedly improved visualization of fine osseous and pulmonary structures, while no consistent qualitative benefit for soft tissue was visually observed. Under matched polychromatic conditions, soft-tissue contrast improvements were modest and tissue-dependent. The most pronounced soft-tissue advantages were observed with spectral imaging, where virtual mono-energetic reconstructions showed a strong energy-dependent behavior. CNR improvements were organ-specific, with clear benefits for renal parenchyma but not for liver tissue. These findings indicate that the principal added value of PCD CT in post-mortem imaging is task-dependent. Detector-related noise reduction provides a robust baseline benefit, high-resolution imaging primarily enhances osseous assessment, and spectral reconstructions offer the greatest potential for soft-tissue optimization.

Forensic ImagingVol. 45
University of Zurich (CH)
Openalex Percentile: Top 20%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.