Irreversible electroporation ablation with bipolar electrodes: ultrasound findings of ablation zones

BACKGROUND: This study aimed to explore the ultrasound and contrast-enhanced ultrasound (CEUS) imaging characteristics and dimensions of ablation lesions after irreversible electroporation of the swine liver and to determine which imaging modality is more suitable for post-procedural follow-up by correlating imaging findings with histopathology. MATERIALS AND METHODS: Irreversible electroporation procedures were conducted on three swine with single bipolar electrodes. All procedures were carried out after laparotomy. Twenty-four ablation zones were created under ultrasound guidance. Ultrasound and CEUS evaluations were performed immediately and 48 h after irreversible electroporation. Liver specimens were harvested 48 h after irreversible electroporation for histopathological analysis. RESULTS: The ablation area appeared as a hypoechoic, well-defined lesion on ultrasound and showed no enhancement on CEUS immediately after irreversible electroporation. At 48 h, the ablation zone appeared as an inhomogeneous hyperechoic area with a hyperechoic margin and blurred boundaries on ultrasound. CEUS clearly delineated the boundary of the ablation zone and demonstrated centripetal enhancement. The dimensions of the ablation area measured on CEUS 48 h after irreversible electroporation showed the highest correlation with the pathologic ablation zone size (length: r = 0.909, width: r = 0.942, p < 0.001), whereas ultrasound measurements showed the lowest correlation (length: r = 0.676, width: r = 0.842, p < 0.001). CONCLUSIONS: Compared with conventional ultrasound, CEUS can accurately measure the dimension of the ablation area, especially 48 h after irreversible electroporation.

Authors

Institutions

Publication Details

Journal
Radiology and Oncology
Published
2026-04-16
DOI
https://doi.org/10.2478/raon-2026-0022
Primary Topic
Microbial Inactivation Methods
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Irreversible electroporation ablation with bipolar electrodes: ultrasound findings of ablation zones

LinYu Zhou, C. Zhang, Shengyong Yin, Ju Li et al.
Radiology and Oncology
Microbial Inactivation Methods
article

Irreversible electroporation ablation with bipolar electrodes: ultrasound findings of ablation zones

LinYu Zhou, C. Zhang, Shengyong Yin, Ju Li, Qiang Chen, Min Xu, Tian’an Jiang
article en

Abstract

BACKGROUND: This study aimed to explore the ultrasound and contrast-enhanced ultrasound (CEUS) imaging characteristics and dimensions of ablation lesions after irreversible electroporation of the swine liver and to determine which imaging modality is more suitable for post-procedural follow-up by correlating imaging findings with histopathology. MATERIALS AND METHODS: Irreversible electroporation procedures were conducted on three swine with single bipolar electrodes. All procedures were carried out after laparotomy. Twenty-four ablation zones were created under ultrasound guidance. Ultrasound and CEUS evaluations were performed immediately and 48 h after irreversible electroporation. Liver specimens were harvested 48 h after irreversible electroporation for histopathological analysis. RESULTS: The ablation area appeared as a hypoechoic, well-defined lesion on ultrasound and showed no enhancement on CEUS immediately after irreversible electroporation. At 48 h, the ablation zone appeared as an inhomogeneous hyperechoic area with a hyperechoic margin and blurred boundaries on ultrasound. CEUS clearly delineated the boundary of the ablation zone and demonstrated centripetal enhancement. The dimensions of the ablation area measured on CEUS 48 h after irreversible electroporation showed the highest correlation with the pathologic ablation zone size (length: r = 0.909, width: r = 0.942, p < 0.001), whereas ultrasound measurements showed the lowest correlation (length: r = 0.676, width: r = 0.842, p < 0.001). CONCLUSIONS: Compared with conventional ultrasound, CEUS can accurately measure the dimension of the ablation area, especially 48 h after irreversible electroporation.

Radiology and Oncology
Zhejiang Academy of Medical Sciences (CN), Ministry of Public Health (MG), First Affiliated Hospital Zhejiang University (CN), Department of Education of Zhejiang Province (CN)
Openalex Percentile: Top 9%
Microbial Inactivation Methods
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.