In Vitro Hemolytic Performance of the Novel Realheart Total Artificial Heart Versus Syn <scp>C</scp> ardia Total Artificial Heart With Human Blood

BACKGROUND: The SynCardia Total Artificial Heart (TAH) is FDA-approved for bridge to transplantation and has been implanted in over 2000 patients. While hemolysis has been reported postimplantation, no published in vitro test data assess hemolysis or hematology. The Realheart TAH is a novel four-chamber pulsatile positive displacement TAH which mimics the natural heart's atrioventricular plane movement for physiological flow patterns and gentle blood handling. This study evaluates in vitro hemolytic performance of the SynCardia TAH against a newer prototype of the Realheart TAH, V11.2, using human blood. METHODS: Hemolysis testing followed ASTM standards, F1830-97 and F1841-97 (2017) using pooled human whole blood diluted to a hematocrit of 30% ± 2%. The targeted flow and pressure head were 5 ± 0.25 L/min and 100 mmHg, respectively. Hematology parameters were analyzed, plasma free hemoglobin (pfHb) was measured, and milligram normalized index of hemolysis (mgNIH) and modified index of hemolysis (MIH) were calculated. Statistical analyses were performed. RESULTS: Mean mgNIH (n = 6) were 18.11 ± 3.53 mg/100 L for V11.2 and 37.15 ± 12.42 mg/100 L for SynCardia (p < 0.05), a 2.05-fold increase in SynCardia TAH. MIH values were 1.69 ± 0.31 for V11.2 and 3.67 ± 1.29 for SynCardia (p < 0.05). Compared with the static control, SynCardia significantly reduced erythrocyte and increased platelet counts, effects not observed with the V11.2, whereas V11.2 significantly reduced leukocyte counts relative to the control. CONCLUSION: Realheart TAH V11.2 exhibited significantly lower hemolysis than SynCardia TAH in a modified systemic circulation test rig, indicating better hemocompatibility.

Authors

Institutions

Publication Details

Journal
Artificial Organs
Published
2026-04-20
DOI
https://doi.org/10.1111/aor.70149
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In Vitro Hemolytic Performance of the Novel Realheart Total Artificial Heart Versus Syn C ardia Total Artificial Heart With Human Blood

Thomas Finocchiaro, Shaikh Faisal Zaman, Oliver Voigt, Kerstin Jönsson Videsäter et al.
Artificial Organs
Mechanical Circulatory Support Devices
article

In Vitro Hemolytic Performance of the Novel Realheart Total Artificial Heart Versus Syn C ardia Total Artificial Heart With Human Blood

Thomas Finocchiaro, Shaikh Faisal Zaman, Oliver Voigt, Kerstin Jönsson Videsäter, Ina Laura Perkins
article en

Abstract

BACKGROUND: The SynCardia Total Artificial Heart (TAH) is FDA-approved for bridge to transplantation and has been implanted in over 2000 patients. While hemolysis has been reported postimplantation, no published in vitro test data assess hemolysis or hematology. The Realheart TAH is a novel four-chamber pulsatile positive displacement TAH which mimics the natural heart's atrioventricular plane movement for physiological flow patterns and gentle blood handling. This study evaluates in vitro hemolytic performance of the SynCardia TAH against a newer prototype of the Realheart TAH, V11.2, using human blood. METHODS: Hemolysis testing followed ASTM standards, F1830-97 and F1841-97 (2017) using pooled human whole blood diluted to a hematocrit of 30% ± 2%. The targeted flow and pressure head were 5 ± 0.25 L/min and 100 mmHg, respectively. Hematology parameters were analyzed, plasma free hemoglobin (pfHb) was measured, and milligram normalized index of hemolysis (mgNIH) and modified index of hemolysis (MIH) were calculated. Statistical analyses were performed. RESULTS: Mean mgNIH (n = 6) were 18.11 ± 3.53 mg/100 L for V11.2 and 37.15 ± 12.42 mg/100 L for SynCardia (p < 0.05), a 2.05-fold increase in SynCardia TAH. MIH values were 1.69 ± 0.31 for V11.2 and 3.67 ± 1.29 for SynCardia (p < 0.05). Compared with the static control, SynCardia significantly reduced erythrocyte and increased platelet counts, effects not observed with the V11.2, whereas V11.2 significantly reduced leukocyte counts relative to the control. CONCLUSION: Realheart TAH V11.2 exhibited significantly lower hemolysis than SynCardia TAH in a modified systemic circulation test rig, indicating better hemocompatibility.

Artificial Organs
Karolinska University Hospital (SE)
Scandinavian Real Heart
Good health and well-being
Openalex Percentile: Top 13%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.