Assessment of Dose Calculation Accuracy for Bone Cement Implants in Radiation Therapy Planning Systems

BACKGROUND AND PURPOSE: This study investigated depth-dose alterations caused by bone cement in solid phantoms during vertebroplasty. Dose variations of megavoltage (MV) photon beams after traversing high-density bone cement were measured to estimate the effective density of bone cement. METHODS: A phantom mimicking human anatomy was fabricated with 2.5-cm acrylic embedded with bone cement. An Elekta VersaHD linear accelerator was used with 6 MV and 6 MV flattening filter-free (FFF) photon beams, field sizes ranging from 3 × 3 cm2 to 10 × 10 cm2. Depth dose, planar dose, and point dose were measured using an Advanced Markus chamber, PTW 1600 two-dimensional array, and Semiflex three-dimensional ion chamber. These were compared with calculated doses from treatment plans with various overridden density values for bone cement. RESULTS: Minimal differences in percent depth dose were observed after MV photons (6 MV and 6 MV FFF) penetrated the bone cement. At 1 cm beyond cement, measured versus calculated doses differed by 1.2% to 2.7%. Within 1 to 5 cm behind cement, dose discrepancies were within 3% to 5%. At 3-cm depth behind cement, point-dose differences measured by the Semiflex chamber indicated smaller deviations at lower override densities, with an optimal match at 0.6 g/cm3. The two-dimensional array showed near-perfect gamma passing rates (∼100%) at an overridden density of 0.8 g/cm3, demonstrating the smallest center-point dose discrepancies. A higher override density of 1.4 g/cm3 yielded results similar to the plan without density override, suggesting minimal advantage at higher override densities. CONCLUSION: This study revealed that dose perturbations induced by bone cement caused the treatment planning system to underestimate actual delivered doses by approximately 3% to 5%. Optimal density override values for bone cement in treatment planning systems appear to be around 0.6 to 0.8 g/cm3. Clinical trials and further data acquisition are necessary to validate precise density values due to variability in cement composition and incomplete cement infusion.

Authors

Institutions

Publication Details

Journal
Journal of the American Academy of Orthopaedic Surgeons
Published
2026-04-21
DOI
https://doi.org/10.5435/jaaos-d-25-00869
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of Dose Calculation Accuracy for Bone Cement Implants in Radiation Therapy Planning Systems

Po-Wei Wu, Wei-Hsun Wang, Yuan-Chun Lai
Journal of the American Academy of Orthopaedic Surgeons
Advanced Radiotherapy Techniques
article

Assessment of Dose Calculation Accuracy for Bone Cement Implants in Radiation Therapy Planning Systems

Po-Wei Wu, Wei-Hsun Wang, Yuan-Chun Lai
article en

Abstract

BACKGROUND AND PURPOSE: This study investigated depth-dose alterations caused by bone cement in solid phantoms during vertebroplasty. Dose variations of megavoltage (MV) photon beams after traversing high-density bone cement were measured to estimate the effective density of bone cement. METHODS: A phantom mimicking human anatomy was fabricated with 2.5-cm acrylic embedded with bone cement. An Elekta VersaHD linear accelerator was used with 6 MV and 6 MV flattening filter-free (FFF) photon beams, field sizes ranging from 3 × 3 cm2 to 10 × 10 cm2. Depth dose, planar dose, and point dose were measured using an Advanced Markus chamber, PTW 1600 two-dimensional array, and Semiflex three-dimensional ion chamber. These were compared with calculated doses from treatment plans with various overridden density values for bone cement. RESULTS: Minimal differences in percent depth dose were observed after MV photons (6 MV and 6 MV FFF) penetrated the bone cement. At 1 cm beyond cement, measured versus calculated doses differed by 1.2% to 2.7%. Within 1 to 5 cm behind cement, dose discrepancies were within 3% to 5%. At 3-cm depth behind cement, point-dose differences measured by the Semiflex chamber indicated smaller deviations at lower override densities, with an optimal match at 0.6 g/cm3. The two-dimensional array showed near-perfect gamma passing rates (∼100%) at an overridden density of 0.8 g/cm3, demonstrating the smallest center-point dose discrepancies. A higher override density of 1.4 g/cm3 yielded results similar to the plan without density override, suggesting minimal advantage at higher override densities. CONCLUSION: This study revealed that dose perturbations induced by bone cement caused the treatment planning system to underestimate actual delivered doses by approximately 3% to 5%. Optimal density override values for bone cement in treatment planning systems appear to be around 0.6 to 0.8 g/cm3. Clinical trials and further data acquisition are necessary to validate precise density values due to variability in cement composition and incomplete cement infusion.

Journal of the American Academy of Orthopaedic Surgeons
Chaoyang University of Technology (TW), National Chung Hsing University (TW), Changhua Christian Hospital (TW), Central Taiwan University of Science and Technology (TW)
Sustainable cities and communities
Openalex Percentile: Top 9%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.