Correcting <scp> <sup>129</sup> Xe </scp> Gas Exchange <scp>MRI</scp> for Incidental Gas‐Phase Excitation—Comparing Approaches, and Identifying Acceptable Thresholds for Reliable Quantification

PURPOSE: Xe gas exchange MRI introduces contamination that can bias membrane uptake and red blood cell (RBC) transfer metrics. Here, we defined the level of gas-phase contamination that is acceptable, and compared the dual-echo (Hahn) and spectroscopy-informed (Willmering) frameworks to correct for it. METHODS: Xe MRI and spectroscopy. The dissolved-phase RF pulse was shortened from its optimal value to increase incidental gas contamination. The Hahn and Willmering corrections were implemented and refined using a grid search over a contamination-dominated ROI to determine any platform-specific scale and phase adjustments. The effects of contamination phase and scale were tested by synthetically reintroducing gas signal at phase offsets that maximized defect- and high intensity-like artifacts. Membrane/gas and RBC/gas defect and high percentages were evaluated against thresholds derived from a healthy reference cohort to determine acceptable gas contamination. RESULTS: Images acquired with the shortened pulse contained 24% ± 6% gas contamination. After applying empirically calibrated phase shifts (Hahn≈230°, Willmering≈49°) and no scale adjustment (optimal value≈1), both correction methods increased RBC SNR by ∼9% and reduced background noise. Phase- and scale-sweep experiments indicated that gas contamination becomes unacceptable once it exceeds ∼9% ± 3%, at which point membrane and RBC defect or high percentages surpass their thresholds. CONCLUSIONS: Gas-phase contamination beyond ∼9% requires correction. After a one-time empirical phase calibration, this can be equally well achieved by both the Hahn and Willmering methods.

Authors

Institutions

Publication Details

Journal
Magnetic Resonance in Medicine
Published
2026-04-20
DOI
https://doi.org/10.1002/mrm.70390
Primary Topic
Atomic and Subatomic Physics Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Correcting 129 Xe Gas Exchange MRI for Incidental Gas‐Phase Excitation—Comparing Approaches, and Identifying Acceptable Thresholds for Reliable Quantification

Andrew D. Hahn, John P. Mugler, Bastiaan Driehuys, Matthew M. Willmering et al.
Magnetic Resonance in Medicine
Atomic and Subatomic Physics Research
article

Correcting 129 Xe Gas Exchange MRI for Incidental Gas‐Phase Excitation—Comparing Approaches, and Identifying Acceptable Thresholds for Reliable Quantification

Andrew D. Hahn, John P. Mugler, Bastiaan Driehuys, Matthew M. Willmering, David Mummy, Suphachart Leewiwatwong
article en

Abstract

PURPOSE: Xe gas exchange MRI introduces contamination that can bias membrane uptake and red blood cell (RBC) transfer metrics. Here, we defined the level of gas-phase contamination that is acceptable, and compared the dual-echo (Hahn) and spectroscopy-informed (Willmering) frameworks to correct for it. METHODS: Xe MRI and spectroscopy. The dissolved-phase RF pulse was shortened from its optimal value to increase incidental gas contamination. The Hahn and Willmering corrections were implemented and refined using a grid search over a contamination-dominated ROI to determine any platform-specific scale and phase adjustments. The effects of contamination phase and scale were tested by synthetically reintroducing gas signal at phase offsets that maximized defect- and high intensity-like artifacts. Membrane/gas and RBC/gas defect and high percentages were evaluated against thresholds derived from a healthy reference cohort to determine acceptable gas contamination. RESULTS: Images acquired with the shortened pulse contained 24% ± 6% gas contamination. After applying empirically calibrated phase shifts (Hahn≈230°, Willmering≈49°) and no scale adjustment (optimal value≈1), both correction methods increased RBC SNR by ∼9% and reduced background noise. Phase- and scale-sweep experiments indicated that gas contamination becomes unacceptable once it exceeds ∼9% ± 3%, at which point membrane and RBC defect or high percentages surpass their thresholds. CONCLUSIONS: Gas-phase contamination beyond ∼9% requires correction. After a one-time empirical phase calibration, this can be equally well achieved by both the Hahn and Willmering methods.

Magnetic Resonance in Medicine
University of Iowa (US), Cincinnati Children's Hospital Medical Center (US), Duke University (US), University of Virginia (US), University of Cincinnati (US)
Openalex Percentile: Top 12%
Atomic and Subatomic Physics Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.