A Portable Fluorometer for the Detection of Glyphosate
periplasmic binding protein PhnD was optimized through strategic fluorophore placement to produce a robust fluorescence signal increase upon glyphosate binding. The biosensor was integrated into a self-contained, 3D-printed device that functions as a miniature fluorometer, providing a simple yes-or-no output for non-expert users while retaining access to raw fluorescence data. The device exhibits nanomolar fluorescence sensitivity with results comparable to a benchtop fluorometer. Using this platform, glyphosate was reliably detected in buffered solutions, commercial herbicides, tap water, and soil extracts. To mitigate false positives arising from phosphate interference, we developed a dual-sensor strategy incorporating an independent phosphate biosensor and a second-generation device capable of multi-wavelength fluorescence detection. Together, these results demonstrate an affordable and versatile biosensing platform with strong potential for field-based environmental monitoring.
Authors
- Shahir S. Rizk (ORCID: https://orcid.org/0009-0009-6996-4098)
- Owen K. Rizk
- Adrian S. Rizk
- Nathanael B. Smith
Institutions
- Indiana University South Bend (US)
Publication Details
- Journal
- Biosensors
- Published
- 2026-04-20
- DOI
- https://doi.org/10.3390/bios16040225
- Primary Topic
- Pesticide and Herbicide Environmental Studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Burroughs Wellcome Fund