Two Cases of Trocar Bending of the Injector During iStent Inject W Implantation

We report two cases of trocar tip bending during implantation of the iStent inject W in combination with cataract surgery. In both cases, the trocar within the insertion tube was bent at a similar location and direction. In Case 1, the bending was recognized after stent deployment, although surgical video review confirmed that it had already been present before deployment. In Case 2, the bending was identified prior to insertion of the injector tip into the anterior chamber. In both cases, the stents initially failed to be properly implanted, and successful implantation was achieved using a new injector. Postoperative courses were uneventful except for transient hyphema. Stereomicroscopic examination revealed that the trocar was bent approximately 0.9-1.0 mm from the tip. These findings suggest that bent-type trocar damage can occur when the exposed trocar tip catches on the corneal stroma during passage through the corneal incision, particularly with the iS2 injector system. Recognition of this damage before stent deployment and prompt replacement of the device are essential to avoid improper implantation and ensure surgical safety.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2026-04-17
DOI
https://doi.org/10.7759/cureus.107218
Primary Topic
Intraocular Surgery and Lenses
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Two Cases of Trocar Bending of the Injector During iStent Inject W Implantation

Sachiko Kaidzu, Yuto Yoshida, Kazunobu Sugihara, Keigo Takagi et al.
Cureus
Intraocular Surgery and Lenses
article

Two Cases of Trocar Bending of the Injector During iStent Inject W Implantation

Sachiko Kaidzu, Yuto Yoshida, Kazunobu Sugihara, Keigo Takagi, Chisako Ida, Hinako Ohtani, Masaki Tanito
article en

Abstract

We report two cases of trocar tip bending during implantation of the iStent inject W in combination with cataract surgery. In both cases, the trocar within the insertion tube was bent at a similar location and direction. In Case 1, the bending was recognized after stent deployment, although surgical video review confirmed that it had already been present before deployment. In Case 2, the bending was identified prior to insertion of the injector tip into the anterior chamber. In both cases, the stents initially failed to be properly implanted, and successful implantation was achieved using a new injector. Postoperative courses were uneventful except for transient hyphema. Stereomicroscopic examination revealed that the trocar was bent approximately 0.9-1.0 mm from the tip. These findings suggest that bent-type trocar damage can occur when the exposed trocar tip catches on the corneal stroma during passage through the corneal incision, particularly with the iS2 injector system. Recognition of this damage before stent deployment and prompt replacement of the device are essential to avoid improper implantation and ensure surgical safety.

CureusVol. 18(4)
Shimane University (JP)
Openalex Percentile: Top 6%
Intraocular Surgery and Lenses
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Two Cases of Trocar Bending of the Injector During iStent Inject W Implantation — Sachiko Kaidzu, Yuto Yoshida, et al. · Cureus (2026) | TGRS Research Map | TGRS