Vital signs measurement using a smartphone camera and a designated application at the Hospital-at-Home setting: a clinical validation study

Abstract Background. Hospital-at-Home (HaH) has become a viable alternative to in-hospital stay, worldwide. Reliable vital signs measurement in this setting is of utmost importance. The validity of vital signs acquisition through the use of cellular applications has not been previously explored. Methods. This was a prospective, controlled clinical trial in the HaH setting. We compared vital signs of HaH patients, sequentially appreciated by a cellular application measurements (CAM) and by conventional measurement (CM) performed by a nurse, in the patients’ homes. Results. Vital signs were compared for 91HaH patients (median age 65 years (IQR 22), with 47.3% females), 77.3% patients were classified as bright/white skin tone. CAM heart rate (HR) was median 75 bpm (IQR 21) vs. 77 bpm (IQR 20) by CM, (p = 0.053); CAM of systolic blood pressure (SBP) was 124 mmHg (IQR 24.5) vs. 116 mmHg (IQR 20) by CM, (p = 0.018); CAM of diastolic blood pressure (DBP) was 75.0 mmHg (IQR 16) vs. 76.0 mmHg (IQR 15) by CM, (p = 0.599); CAM respiratory rate (RR) was 18 breaths/min (IQR 6) vs. 16 breaths/min (IQR 5.75) by CM, (p = 0.003); CAM oxygen blood saturation was 97.0% (IQR 3) vs. 98.0% (IQR 1) by CM, (p = 0.002). Agreement between methods measured by Interrater reliability (ICC) was moderate for HR (ICC = 0.71), lower for RR (ICC = 0.37) and poor for SBP measurement (ICC = 0.21). ICC values for DBP and SpO 2 were − 0.03 and 0.06 respectively. In a multivariable analysis, darker skin categories were associated with substantial bias in diastolic blood pressure estimation. Conclusions. Our findings indicate that CAM technology can estimate some physiological parameters, particularly heart rate, but other measurements show limited agreement with standard monitoring. While promising for remote care such as HaH when conventional measurement is unavailable, broader validation, algorithm improvement, and rigorous clinical evaluation are essential before routine, widespread clinical use.

Authors

Institutions

Publication Details

Journal
BioMedical Engineering OnLine
Published
2026-11-07
DOI
https://doi.org/10.1186/s12938-026-01654-y
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Vital signs measurement using a smartphone camera and a designated application at the Hospital-at-Home setting: a clinical validation study

Anat Ekka Zohar, Gad Segal, Galia Barkai, LIAT AROSHASS et al.
1 links
BioMedical Engineering OnLine
Non-Invasive Vital Sign Monitoring
article

Vital signs measurement using a smartphone camera and a designated application at the Hospital-at-Home setting: a clinical validation study

Anat Ekka Zohar, Gad Segal, Galia Barkai, LIAT AROSHASS, Boris Fizdel, Hila Hakim, Noi Meersohn, Rachel Sarafraz, Dan Rozenberg, Or Dagan
article en

Abstract

Abstract Background. Hospital-at-Home (HaH) has become a viable alternative to in-hospital stay, worldwide. Reliable vital signs measurement in this setting is of utmost importance. The validity of vital signs acquisition through the use of cellular applications has not been previously explored. Methods. This was a prospective, controlled clinical trial in the HaH setting. We compared vital signs of HaH patients, sequentially appreciated by a cellular application measurements (CAM) and by conventional measurement (CM) performed by a nurse, in the patients’ homes. Results. Vital signs were compared for 91HaH patients (median age 65 years (IQR 22), with 47.3% females), 77.3% patients were classified as bright/white skin tone. CAM heart rate (HR) was median 75 bpm (IQR 21) vs. 77 bpm (IQR 20) by CM, (p = 0.053); CAM of systolic blood pressure (SBP) was 124 mmHg (IQR 24.5) vs. 116 mmHg (IQR 20) by CM, (p = 0.018); CAM of diastolic blood pressure (DBP) was 75.0 mmHg (IQR 16) vs. 76.0 mmHg (IQR 15) by CM, (p = 0.599); CAM respiratory rate (RR) was 18 breaths/min (IQR 6) vs. 16 breaths/min (IQR 5.75) by CM, (p = 0.003); CAM oxygen blood saturation was 97.0% (IQR 3) vs. 98.0% (IQR 1) by CM, (p = 0.002). Agreement between methods measured by Interrater reliability (ICC) was moderate for HR (ICC = 0.71), lower for RR (ICC = 0.37) and poor for SBP measurement (ICC = 0.21). ICC values for DBP and SpO 2 were − 0.03 and 0.06 respectively. In a multivariable analysis, darker skin categories were associated with substantial bias in diastolic blood pressure estimation. Conclusions. Our findings indicate that CAM technology can estimate some physiological parameters, particularly heart rate, but other measurements show limited agreement with standard monitoring. While promising for remote care such as HaH when conventional measurement is unavailable, broader validation, algorithm improvement, and rigorous clinical evaluation are essential before routine, widespread clinical use.

BioMedical Engineering OnLine
Tel Aviv University (IL), University of Nicosia (CY), Sheba Medical Center (IL)
Good health and well-being
Openalex Percentile: Top 85%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon