Comparison of intraocular pressure measured by Goldman applanation tonometer, noncontact tonometer and trans-palpebral tonometer

To evaluate reliability and agreement of non-contact tonometry (NCT) and trans-palpebral tonometry (TPT) compared with Goldmann-applanation tonometry (GAT). A total of 224 eyes from 112 patients were included between January and April 2025. Intraocular pressure (IOP) was measured using GAT, NCT, and TPT. Comparison among the three measurement groups were performed using the related-samples Friedman test. Post hoc pairwise comparisons were conducted using the Wilcoxon signed-rank test with Bonferroni correction. Reliability was assessed using Cronbach’s alpha and the intraclass correlation coefficient (ICC). Agreement between methods was evaluated using Bland–Altman plots and Spearman’s rank-order correlation. Median IOP were 14 mmHg for GAT, 14 mmHg for NCT and 15 mmHg for TPT. Friedman’s ANOVA showed a statistically significant difference among three methods (χ² = 40.652, p < 0.001). TPT readings differed significantly from both GAT and NCT (p < 0.001). The difference between GAT and NCT was not significant (p = 0.024) because it did not meet the Bonferroni-adjusted threshold (α = 0.017). Cronbach’s alpha was 0.691, indicating acceptable internal consistency. The ICC for average IOP was 0.668 (95% CI: 0.571–0.743, p < 0.001), suggesting moderate reliability. Spearman’s correlation showed moderate-agreement between GAT and NCT (ρ = 0.586, p < 0.001), weak-agreement between TPT and NCT (ρ = 0.315, p < 0.001), and no correlation between GAT and TPT (ρ = 0.090, p = 0.177). Bland-Altman analysis confirmed good-agreement between GAT and NCT, fair-agreement between NCT and TPT, and poor-agreement between GAT and TPT. GAT and NCT showed comparatively better agreement, whereas TPT demonstrated wide limits of agreement with GAT and should not be used interchangeably for clinical decision-making; TPT may be considered only for preliminary screening when contact tonometry is not feasible.

Authors

Institutions

Publication Details

Journal
BMC Ophthalmology
Published
2026-04-17
DOI
https://doi.org/10.1186/s12886-026-04821-w
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of intraocular pressure measured by Goldman applanation tonometer, noncontact tonometer and trans-palpebral tonometer

Madhu Thapa, G. C. Saroj, Gautam Pragati Adhikari, Gautam Kumar
BMC Ophthalmology
Corneal surgery and disorders
article

Comparison of intraocular pressure measured by Goldman applanation tonometer, noncontact tonometer and trans-palpebral tonometer

Madhu Thapa, G. C. Saroj, Gautam Pragati Adhikari, Gautam Kumar
article en

Abstract

To evaluate reliability and agreement of non-contact tonometry (NCT) and trans-palpebral tonometry (TPT) compared with Goldmann-applanation tonometry (GAT). A total of 224 eyes from 112 patients were included between January and April 2025. Intraocular pressure (IOP) was measured using GAT, NCT, and TPT. Comparison among the three measurement groups were performed using the related-samples Friedman test. Post hoc pairwise comparisons were conducted using the Wilcoxon signed-rank test with Bonferroni correction. Reliability was assessed using Cronbach’s alpha and the intraclass correlation coefficient (ICC). Agreement between methods was evaluated using Bland–Altman plots and Spearman’s rank-order correlation. Median IOP were 14 mmHg for GAT, 14 mmHg for NCT and 15 mmHg for TPT. Friedman’s ANOVA showed a statistically significant difference among three methods (χ² = 40.652, p < 0.001). TPT readings differed significantly from both GAT and NCT (p < 0.001). The difference between GAT and NCT was not significant (p = 0.024) because it did not meet the Bonferroni-adjusted threshold (α = 0.017). Cronbach’s alpha was 0.691, indicating acceptable internal consistency. The ICC for average IOP was 0.668 (95% CI: 0.571–0.743, p < 0.001), suggesting moderate reliability. Spearman’s correlation showed moderate-agreement between GAT and NCT (ρ = 0.586, p < 0.001), weak-agreement between TPT and NCT (ρ = 0.315, p < 0.001), and no correlation between GAT and TPT (ρ = 0.090, p = 0.177). Bland-Altman analysis confirmed good-agreement between GAT and NCT, fair-agreement between NCT and TPT, and poor-agreement between GAT and TPT. GAT and NCT showed comparatively better agreement, whereas TPT demonstrated wide limits of agreement with GAT and should not be used interchangeably for clinical decision-making; TPT may be considered only for preliminary screening when contact tonometry is not feasible.

BMC OphthalmologyVol. 26(1)
Tribhuvan University Teaching Hospital (NP), Primary Health Care (QA)
No poverty
Openalex Percentile: Top 18%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.