Rapid fluorescence excitation-emission matrix measurements using hybrid configuration of Michelson interferometer and grating spectrometer

A technique for rapid fluorescence excitation-emission matrix (EEM) measurements is proposed, which leverages the high optical throughput and tunable resolution of a scanning Michelson interferometer and the fast acquisition of a grating spectrometer. In this technique, a broadband excitation light beam is modulated by a galvanometer-scanning Michelson interferometer, and the resulting interferometrically modulated light excites the sample. By synchronizing fluorescence detection with galvanometer scanning, fluorescence spectra corresponding to different optical path differences are separated. This configuration enables the tradeoff between the excitation module’s spectral resolution and measurement speed, while enabling high throughput excitation light focused into a small spot on the sample. It is demonstrated that with excitation and emission spectral resolutions of 10 nm and 0.44 nm, respectively, the system can acquire the EEM of an 8.34 μM Rhodamine 6 G solution in ethanol in only 977 ms.

Authors

Publication Details

Journal
Optics Letters
Published
2026-04-17
DOI
https://doi.org/10.1364/ol.589714
Primary Topic
Optical Polarization and Ellipsometry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Rapid fluorescence excitation-emission matrix measurements using hybrid configuration of Michelson interferometer and grating spectrometer

Fuzhou Shen, Xingyun Wang, Quan Liu, JUNJIANG LIU et al.
Optics Letters
Optical Polarization and Ellipsometry
article

Rapid fluorescence excitation-emission matrix measurements using hybrid configuration of Michelson interferometer and grating spectrometer

Fuzhou Shen, Xingyun Wang, Quan Liu, JUNJIANG LIU, Xiangguo Yuan, Jisheng Zhang
article en

Abstract

A technique for rapid fluorescence excitation-emission matrix (EEM) measurements is proposed, which leverages the high optical throughput and tunable resolution of a scanning Michelson interferometer and the fast acquisition of a grating spectrometer. In this technique, a broadband excitation light beam is modulated by a galvanometer-scanning Michelson interferometer, and the resulting interferometrically modulated light excites the sample. By synchronizing fluorescence detection with galvanometer scanning, fluorescence spectra corresponding to different optical path differences are separated. This configuration enables the tradeoff between the excitation module’s spectral resolution and measurement speed, while enabling high throughput excitation light focused into a small spot on the sample. It is demonstrated that with excitation and emission spectral resolutions of 10 nm and 0.44 nm, respectively, the system can acquire the EEM of an 8.34 μM Rhodamine 6 G solution in ethanol in only 977 ms.

Optics LettersVol. 51(10)
Affordable and clean energy
Openalex Percentile: Top 25%
Optical Polarization and Ellipsometry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.