Abstract Background Radiometabolic 177 Lu-based treatment is a process that involves different specialists. Due to its high complexity and to the potential serious risk of health damages for patients and healthcare professionals, a global analysis of the whole process would allow to identify the weak points and then to improve clinical standards. Radioprotection risk assessment also fulfils the requirement of the 2013/59/EURATOM directive. Aim To identify the main critical points in the process of 177 Lu therapies by using the failure mode and effects analysis (FMEA) method. The focus of the FMEA analysis is the risk to the patient. Materials and methods We divided the whole process of 177 Lu-based therapies into main steps and identified all possible failure modes. Each failure mode was scored for severity (S), occurrence (O) and detectability (D), range 1–10. Based on the Risk Priority Number (RPN = SxOxD), all failures were ranked. Both in-house labelled or already and ready-for-injection radiopharmaceuticals were considered, as well as both 177 Lu-PSMA and 177 Lu-DOTATATE/TOC therapies. Results The failure modes with the highest RPN scores identified were: incorrect infusion supervision based on count measurements, incorrect catheter connections (vial–injection pump–patient), incorrect end of infusion, and false-negative pregnancy test results. Additionally, for in-house labelled radiopharmaceuticals, the incorrect selection of calibration factors in the activimetre for vial activity measure was identified as a critical issue. Conclusions The FMEA analysis allowed the identification of the main failure mode in 177 Lu treatment procedures. Based on the obtained results, strategies could be implemented in clinical practice in order to further improve patient safety. The present results may help the new implementation of 177 Lu-based therapy in the perspective of increasing use of these therapies.

Authors

Institutions

Publication Details

Journal
The European Physical Journal Plus
Published
2026-04-16
DOI
https://doi.org/10.1140/epjp/s13360-026-07582-z
Primary Topic
Radiopharmaceutical Chemistry and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

From directive to practice: applying FMEA to 177Lu therapy according to Euratom 2013/59

Irene Marini, Valentina Di Iorio, M. Rustignoli, M. Golinucci et al.
The European Physical Journal Plus
Radiopharmaceutical Chemistry and Applications
article

From directive to practice: applying FMEA to 177Lu therapy according to Euratom 2013/59

Irene Marini, Valentina Di Iorio, M. Rustignoli, M. Golinucci, Maria Luisa Belli, Anna Sarnelli, Federica Matteucci, Anita Zeneli, Elena Prati, Maria Teresa Montella, Gianina Maria Volpe
article en

Abstract

Abstract Background Radiometabolic 177 Lu-based treatment is a process that involves different specialists. Due to its high complexity and to the potential serious risk of health damages for patients and healthcare professionals, a global analysis of the whole process would allow to identify the weak points and then to improve clinical standards. Radioprotection risk assessment also fulfils the requirement of the 2013/59/EURATOM directive. Aim To identify the main critical points in the process of 177 Lu therapies by using the failure mode and effects analysis (FMEA) method. The focus of the FMEA analysis is the risk to the patient. Materials and methods We divided the whole process of 177 Lu-based therapies into main steps and identified all possible failure modes. Each failure mode was scored for severity (S), occurrence (O) and detectability (D), range 1–10. Based on the Risk Priority Number (RPN = SxOxD), all failures were ranked. Both in-house labelled or already and ready-for-injection radiopharmaceuticals were considered, as well as both 177 Lu-PSMA and 177 Lu-DOTATATE/TOC therapies. Results The failure modes with the highest RPN scores identified were: incorrect infusion supervision based on count measurements, incorrect catheter connections (vial–injection pump–patient), incorrect end of infusion, and false-negative pregnancy test results. Additionally, for in-house labelled radiopharmaceuticals, the incorrect selection of calibration factors in the activimetre for vial activity measure was identified as a critical issue. Conclusions The FMEA analysis allowed the identification of the main failure mode in 177 Lu treatment procedures. Based on the obtained results, strategies could be implemented in clinical practice in order to further improve patient safety. The present results may help the new implementation of 177 Lu-based therapy in the perspective of increasing use of these therapies.

The European Physical Journal PlusVol. 141(4)
Istituto Scientifico Romagnolo per lo Studio e la Cura dei Tumori (IT)
Openalex Percentile: Top 14%
Radiopharmaceutical Chemistry and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.