Regression Using 20 Samples to Harmonize a Point‐of‐Care Analyzer With a Commercial Laboratory Analyzer for Feline Plasma Biochemistry Testing

BACKGROUND: Harmonization enables direct comparison of patient results from different analyzers and/or measurement techniques and the use of common reference intervals and decision thresholds. Point-of-care analyzers (POC) cannot be calibrated by the end user, so regression offers an opportunity for calculated harmonization. This approach has been used to harmonize (POC) with commercial laboratory analyzers (CL) with > 90 paired feline plasma samples. OBJECTIVE: Our aims were to use 20 pairs of results to derive regression equations to harmonize the POC results with those of the CL. Harmonized results were assessed by recognized quality goals. METHODS: Systematic sampling was used to choose the 20 samples used to calculate regression parameters (with three techniques). These were applied to the remainder of the original dataset to harmonize the POC results to those from the CL. POC results were assessed for bias and the number of results reaching quality goals before and after adjustment. RESULTS: Regression from 20 paired samples was able to harmonize POC analyzer results to a commercial laboratory analyzer for 16 feline plasma biochemistry measurands. CONCLUSIONS: Most analytes could be harmonized by this approach. Reasons for harmonization failure were inadequate concentration range of patient results (AST, lipase, total bilirubin, GGT), potential non-linear relationship (lipase), and no clinical performance goal (globulins). Additional studies using samples with clinically important values may overcome these limitations, and similar studies of other species and other types of point-of-care analyzers would be of interest.

Authors

Institutions

Publication Details

Journal
Veterinary Clinical Pathology
Published
2026-04-15
DOI
https://doi.org/10.1111/vcp.70108
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Regression Using 20 Samples to Harmonize a Point‐of‐Care Analyzer With a Commercial Laboratory Analyzer for Feline Plasma Biochemistry Testing

Douglas Hayward, Kathleen P. Freeman, Susan M. Jaensch, Randolph M. Baral et al.
Veterinary Clinical Pathology
Clinical Laboratory Practices and Quality Control
article

Regression Using 20 Samples to Harmonize a Point‐of‐Care Analyzer With a Commercial Laboratory Analyzer for Feline Plasma Biochemistry Testing

Douglas Hayward, Kathleen P. Freeman, Susan M. Jaensch, Randolph M. Baral, Bente Flatland
article en

Abstract

BACKGROUND: Harmonization enables direct comparison of patient results from different analyzers and/or measurement techniques and the use of common reference intervals and decision thresholds. Point-of-care analyzers (POC) cannot be calibrated by the end user, so regression offers an opportunity for calculated harmonization. This approach has been used to harmonize (POC) with commercial laboratory analyzers (CL) with > 90 paired feline plasma samples. OBJECTIVE: Our aims were to use 20 pairs of results to derive regression equations to harmonize the POC results with those of the CL. Harmonized results were assessed by recognized quality goals. METHODS: Systematic sampling was used to choose the 20 samples used to calculate regression parameters (with three techniques). These were applied to the remainder of the original dataset to harmonize the POC results to those from the CL. POC results were assessed for bias and the number of results reaching quality goals before and after adjustment. RESULTS: Regression from 20 paired samples was able to harmonize POC analyzer results to a commercial laboratory analyzer for 16 feline plasma biochemistry measurands. CONCLUSIONS: Most analytes could be harmonized by this approach. Reasons for harmonization failure were inadequate concentration range of patient results (AST, lipase, total bilirubin, GGT), potential non-linear relationship (lipase), and no clinical performance goal (globulins). Additional studies using samples with clinically important values may overcome these limitations, and similar studies of other species and other types of point-of-care analyzers would be of interest.

Veterinary Clinical Pathology
The University of Sydney (AU), Paddington Cat Hospital (AU), University of Tennessee at Knoxville (US)
Openalex Percentile: Top 14%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.