Comparative Performance, Procedural Characteristics and Safety of Micra Versus Aveir Leadless Pacemakers: Systematic Review

BACKGROUND: Leadless pacemakers (LPs) are developed to avoid pocket- and lead-related complications of transvenous pacing systems, but direct comparative data between two single chamber LP, namely Micra VR and Aveir VR are scarce. METHODS: A systematic review and meta-analysis of studies directly comparing Micra VR and Aveir VR was performed. PubMed, Embase, and Scopus were searched through November 2025. Eligible studies reported electrical performance, procedural metrics, or periprocedural complications. RESULTS: Four observational studies, including 757 patients (363 Aveir, 394 Micra), met the inclusion criteria, with follow-up of 1-6 months. At implantation, Aveir showed slightly higher pacing thresholds (mean difference (MD) 0.18 V; 95% CI 0.09-0.26), while impedance and R-wave amplitude were similar. In follow-up, pacing thresholds showed no significant difference between devices, whereas Micra demonstrated higher R-wave amplitudes (MD -1.18 mV; 95% CI -2.29 to -0.07). Fluoroscopy time was longer with Aveir (MD 2.89 min; 95% CI 0.69-5.08), but total procedure duration did not differ significantly. Rates of device dislodgement, pericardial effusion/tamponade, and overall periprocedural complications were low and not statistically different between the two systems. CONCLUSION: Micra VR and Aveir VR demonstrate generally comparable short-term electrical performance, procedural efficacy, and safety. Minor differences in thresholds, sensing, and fluoroscopy time are likely indicative of design and workflow characteristics. Larger, well-controlled randomized studies with extended follow-up periods are necessary to establish long-term performance.

Authors

Institutions

Publication Details

Journal
Pacing and Clinical Electrophysiology
Published
2026-04-14
DOI
https://doi.org/10.1111/pace.70253
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Performance, Procedural Characteristics and Safety of Micra Versus Aveir Leadless Pacemakers: Systematic Review

Hamidreza Soleimani, Negin Mohammadi, Reza Mollazadeh, Parham Dastjerdi et al.
Pacing and Clinical Electrophysiology
Cardiac pacing and defibrillation studies
article

Comparative Performance, Procedural Characteristics and Safety of Micra Versus Aveir Leadless Pacemakers: Systematic Review

Hamidreza Soleimani, Negin Mohammadi, Reza Mollazadeh, Parham Dastjerdi, Hedieh Shayesteh, Masoud Eslami, Cyrus Hadadi, Iman Razeghian
article en

Abstract

BACKGROUND: Leadless pacemakers (LPs) are developed to avoid pocket- and lead-related complications of transvenous pacing systems, but direct comparative data between two single chamber LP, namely Micra VR and Aveir VR are scarce. METHODS: A systematic review and meta-analysis of studies directly comparing Micra VR and Aveir VR was performed. PubMed, Embase, and Scopus were searched through November 2025. Eligible studies reported electrical performance, procedural metrics, or periprocedural complications. RESULTS: Four observational studies, including 757 patients (363 Aveir, 394 Micra), met the inclusion criteria, with follow-up of 1-6 months. At implantation, Aveir showed slightly higher pacing thresholds (mean difference (MD) 0.18 V; 95% CI 0.09-0.26), while impedance and R-wave amplitude were similar. In follow-up, pacing thresholds showed no significant difference between devices, whereas Micra demonstrated higher R-wave amplitudes (MD -1.18 mV; 95% CI -2.29 to -0.07). Fluoroscopy time was longer with Aveir (MD 2.89 min; 95% CI 0.69-5.08), but total procedure duration did not differ significantly. Rates of device dislodgement, pericardial effusion/tamponade, and overall periprocedural complications were low and not statistically different between the two systems. CONCLUSION: Micra VR and Aveir VR demonstrate generally comparable short-term electrical performance, procedural efficacy, and safety. Minor differences in thresholds, sensing, and fluoroscopy time are likely indicative of design and workflow characteristics. Larger, well-controlled randomized studies with extended follow-up periods are necessary to establish long-term performance.

Pacing and Clinical Electrophysiology
Imam Khomeini Hospital (IR), MedStar Heart & Vascular Institute (US), Shahid Beheshti University of Medical Sciences (IR), Tehran University of Medical Sciences (IR)
Openalex Percentile: Top 18%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.