Intraprosthetic Dislocation Following Dual-Mobility Total Hip Arthroplasty: A Case Report

Dual-mobility total hip arthroplasty has been increasingly adopted to improve joint stability and reduce postoperative dislocation. Despite these advantages, this implant design is associated with specific modes of failure. One such complication is intraprosthetic dislocation, which occurs when the femoral head disengages from the polyethylene liner. Although uncommon, this condition is clinically important and may not be readily identified at initial presentation. We report a case involving a 67-year-old woman who presented with sudden onset of hip pain and mechanical restriction of movement two years after undergoing dual-mobility total hip arthroplasty for a fracture neck of femur. Radiographic assessment revealed abnormal positioning of the femoral head within the acetabular component, suggesting dissociation of the polyethylene liner. Computed tomography further supported this diagnosis, and inflammatory markers were within normal limits. Surgical exploration confirmed intraprosthetic dislocation of the liner, while both the femoral stem and acetabular shell were well fixed. Management consisted of removing the mobile bearing and converting the articulation to a fixed polyethylene liner within the existing acetabular cup. Postoperatively, the patient recovered without complications and demonstrated sustained functional improvement without recurrent instability at one year. This case underscores that intraprosthetic dislocation remains a potential cause of mechanical failure even with contemporary dual-mobility implants. Because clinical symptoms may be subtle, careful attention to characteristic imaging findings is essential to avoid diagnostic delay. When implant fixation is preserved, selective exchange of modular components can provide a successful and less invasive treatment option. Prompt recognition and appropriate surgical intervention are crucial to prevent further prosthetic damage and to achieve satisfactory functional outcomes.

Authors

Publication Details

Journal
Cureus
Published
2026-04-14
DOI
https://doi.org/10.7759/cureus.107069
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intraprosthetic Dislocation Following Dual-Mobility Total Hip Arthroplasty: A Case Report

Akshay Nayak, Abhinandan Punit, Abin Mahmood Nizar, Yathisha SR et al.
Cureus
Orthopaedic implants and arthroplasty
article

Intraprosthetic Dislocation Following Dual-Mobility Total Hip Arthroplasty: A Case Report

Akshay Nayak, Abhinandan Punit, Abin Mahmood Nizar, Yathisha SR, Yeshwanth M Reddy
article en

Abstract

Dual-mobility total hip arthroplasty has been increasingly adopted to improve joint stability and reduce postoperative dislocation. Despite these advantages, this implant design is associated with specific modes of failure. One such complication is intraprosthetic dislocation, which occurs when the femoral head disengages from the polyethylene liner. Although uncommon, this condition is clinically important and may not be readily identified at initial presentation. We report a case involving a 67-year-old woman who presented with sudden onset of hip pain and mechanical restriction of movement two years after undergoing dual-mobility total hip arthroplasty for a fracture neck of femur. Radiographic assessment revealed abnormal positioning of the femoral head within the acetabular component, suggesting dissociation of the polyethylene liner. Computed tomography further supported this diagnosis, and inflammatory markers were within normal limits. Surgical exploration confirmed intraprosthetic dislocation of the liner, while both the femoral stem and acetabular shell were well fixed. Management consisted of removing the mobile bearing and converting the articulation to a fixed polyethylene liner within the existing acetabular cup. Postoperatively, the patient recovered without complications and demonstrated sustained functional improvement without recurrent instability at one year. This case underscores that intraprosthetic dislocation remains a potential cause of mechanical failure even with contemporary dual-mobility implants. Because clinical symptoms may be subtle, careful attention to characteristic imaging findings is essential to avoid diagnostic delay. When implant fixation is preserved, selective exchange of modular components can provide a successful and less invasive treatment option. Prompt recognition and appropriate surgical intervention are crucial to prevent further prosthetic damage and to achieve satisfactory functional outcomes.

Cureus
Good health and well-being
Openalex Percentile: Top 7%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Intraprosthetic Dislocation Following Dual-Mobility Total Hip Arthroplasty: A Case Report — Akshay Nayak, Abhinandan Punit, et al. · Cureus (2026) | TGRS Research Map | TGRS