Validation of the measurement accuracy of the CONTEC ABPM50 ambulatory blood pressure monitor according to the International Protocol of the European Society of Hypertension ESH-IP 2010 protocol in a resource-limited healthcare setting

The aim of the present study was to validate the ambulatory blood pressure monitoring device CONTEC ABPM50 according to the International Protocol revision 2010 of the European Society of Hypertension (ESH-IP 2010) in a Venezuelan adult population. In 39 adult participants aged 25–82 years, blood pressure measurements were performed alternately via the CONTEC ABPM50 and a reference mercury sphygmomanometer according to the ESH-IP 2010 guidelines. A total of 117 paired comparisons were included for analysis. The measurements covered all required blood pressure ranges and were performed by two blinded observers. The absolute differences between the test device and the reference were within 5 mmHg in 114 of 117 comparisons (97.4%) for systolic blood pressure (SBP) and 116 of 117 (99.1%) for diastolic blood pressure (DBP). The mean ± SD difference between CONTEC ABPM50 and the mercury sphygmomanometer was 0.24 ± 3.54 mmHg for SBP and − 0.85 ± 3.32 mmHg for DBP. Bland–Altman analysis showed a mean bias of − 0.18 mmHg for systolic blood pressure (SBP) with 95% limits of agreement from − 5.96 to + 4.60 mmHg, and a mean bias of + 1.17 mmHg for diastolic blood pressure (DBP) with 95% limits of agreement from − 3.69 to + 6.04 mmHg. The device met all the criteria specified in parts 1 and 2 of the ESH-IP 2010 protocol. The CONTEC ABPM50 met the static accuracy requirements of the ESH-IP 2010. While this study validates the device’s measurement module under controlled conditions, further research is needed to evaluate its performance during dynamic ambulatory activities and nocturnal periods.

Authors

Institutions

Publication Details

Journal
Discover Medicine
Published
2026-04-15
DOI
https://doi.org/10.1007/s44337-026-00608-6
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of the measurement accuracy of the CONTEC ABPM50 ambulatory blood pressure monitor according to the International Protocol of the European Society of Hypertension ESH-IP 2010 protocol in a resource-limited healthcare setting

José Ayala-Hernández, Steffany Castro, Guillermo Cerquoni, Paola Brigli et al.
Discover Medicine
Blood Pressure and Hypertension Studies
article

Validation of the measurement accuracy of the CONTEC ABPM50 ambulatory blood pressure monitor according to the International Protocol of the European Society of Hypertension ESH-IP 2010 protocol in a resource-limited healthcare setting

José Ayala-Hernández, Steffany Castro, Guillermo Cerquoni, Paola Brigli, Cristina López-Sánchez, Ana Barbosa
article en

Abstract

The aim of the present study was to validate the ambulatory blood pressure monitoring device CONTEC ABPM50 according to the International Protocol revision 2010 of the European Society of Hypertension (ESH-IP 2010) in a Venezuelan adult population. In 39 adult participants aged 25–82 years, blood pressure measurements were performed alternately via the CONTEC ABPM50 and a reference mercury sphygmomanometer according to the ESH-IP 2010 guidelines. A total of 117 paired comparisons were included for analysis. The measurements covered all required blood pressure ranges and were performed by two blinded observers. The absolute differences between the test device and the reference were within 5 mmHg in 114 of 117 comparisons (97.4%) for systolic blood pressure (SBP) and 116 of 117 (99.1%) for diastolic blood pressure (DBP). The mean ± SD difference between CONTEC ABPM50 and the mercury sphygmomanometer was 0.24 ± 3.54 mmHg for SBP and − 0.85 ± 3.32 mmHg for DBP. Bland–Altman analysis showed a mean bias of − 0.18 mmHg for systolic blood pressure (SBP) with 95% limits of agreement from − 5.96 to + 4.60 mmHg, and a mean bias of + 1.17 mmHg for diastolic blood pressure (DBP) with 95% limits of agreement from − 3.69 to + 6.04 mmHg. The device met all the criteria specified in parts 1 and 2 of the ESH-IP 2010 protocol. The CONTEC ABPM50 met the static accuracy requirements of the ESH-IP 2010. While this study validates the device’s measurement module under controlled conditions, further research is needed to evaluate its performance during dynamic ambulatory activities and nocturnal periods.

Discover MedicineVol. 3(1)
Hospital Clínico Universitario de Caracas (VE), Universidad Central de Venezuela (VE)
Partnerships for the goals
Openalex Percentile: Top 25%
Blood Pressure and Hypertension Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Validation of the measurement accuracy of the CONTEC ABPM50 ambulatory blood pressure monitor according to the International Protocol of the European Society of Hypertension ESH-IP 2010 protocol in a resource-limited healthcare setting — José Ayala-Hernández, Steffany Castro, et al. · Discover Medicine (2026) | TGRS Research Map | TGRS