Characterization of a successful transseptal access with an electrified guidewire: An ex vivo ovine study

BACKGROUND: The determinants of successful transseptal puncture (TSP) using an electrified guidewire (ETSP) have not been studied. OBJECTIVE: We used an ex vivo ovine model to identify the determinants for successful ETSP. METHODS: 25 fresh ovine hearts were dissected, and lateral right atrial/appendage tissue was excised to create an interatrial septum (IAS) model. The sheath and guidewire were fixed to a custom-made jig, which allowed linear displacement to model clinical catheter manipulation. We observed TSP at varying degrees of IAS displacement (0, 5, 10, and 15 mm) and guidewire tip protrusion (1, 3, and 5 mm). Electrocautery was applied to the proximal end of the guidewire at 15 W in the "Coag" mode for 2 seconds to attempt ETSP. The thermal profiles of the various configurations during electrification were studied using thermal imaging. Finally, thermal assessment of ETSP was studied by comparing 2 common waveforms ("Coag" and "Cut") at 15 W. RESULTS: A total of 58 ETSP runs were performed. The model with a 1-mm wire protrusion and an IAS displacement of 10 mm yielded a probability of 100% to achieve TSP with a single 2-second application. Extended guidewire protrusion (3 and 5 mm) and no IAS displacement resulted in unsuccessful puncture and increased the risk of high tissue temperatures and larger thermal lesions. The "Coag" waveform produced lower peak temperature (34.36°C ± 3.17°C vs 68.6°C ± 3.7°C; P < .0001) and smaller thermal footprint (2.9 ± 0.2 mm vs 5.1 ± 0.2 mm; P < .0001), while maintaining perforation success compared with the "Cut" waveform. CONCLUSION: Displacement of the IAS is required for successful ETSP. Guidewire protrusion beyond 1 mm resulted in inconsistent perforation and increased tissue temperature.

Authors

Institutions

Publication Details

Journal
Heart Rhythm
Published
2026-04-01
DOI
https://doi.org/10.1016/j.hrthm.2026.01.052
Citations
1
Primary Topic
Thyroid and Parathyroid Surgery
Type
article
Field-Weighted Citation Impact
26.17

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterization of a successful transseptal access with an electrified guidewire: An ex vivo ovine study

Alessandra Pina, Alexis Puvrez, N. Weerasooriya, Darius Chapman et al.
1 citations
Heart Rhythm
Thyroid and Parathyroid Surgery
26.17
article

Characterization of a successful transseptal access with an electrified guidewire: An ex vivo ovine study

Alessandra Pina, Alexis Puvrez, N. Weerasooriya, Darius Chapman, C E Wong, Sanjai Pattu Valappil, Prashanthan Sanders, J P Ariyaratnam, Elnaz Shahmohamadi, Adrian Elliott, Samuel J. Tu, Anand N. Ganesan
article en
1 citations

Abstract

BACKGROUND: The determinants of successful transseptal puncture (TSP) using an electrified guidewire (ETSP) have not been studied. OBJECTIVE: We used an ex vivo ovine model to identify the determinants for successful ETSP. METHODS: 25 fresh ovine hearts were dissected, and lateral right atrial/appendage tissue was excised to create an interatrial septum (IAS) model. The sheath and guidewire were fixed to a custom-made jig, which allowed linear displacement to model clinical catheter manipulation. We observed TSP at varying degrees of IAS displacement (0, 5, 10, and 15 mm) and guidewire tip protrusion (1, 3, and 5 mm). Electrocautery was applied to the proximal end of the guidewire at 15 W in the "Coag" mode for 2 seconds to attempt ETSP. The thermal profiles of the various configurations during electrification were studied using thermal imaging. Finally, thermal assessment of ETSP was studied by comparing 2 common waveforms ("Coag" and "Cut") at 15 W. RESULTS: A total of 58 ETSP runs were performed. The model with a 1-mm wire protrusion and an IAS displacement of 10 mm yielded a probability of 100% to achieve TSP with a single 2-second application. Extended guidewire protrusion (3 and 5 mm) and no IAS displacement resulted in unsuccessful puncture and increased the risk of high tissue temperatures and larger thermal lesions. The "Coag" waveform produced lower peak temperature (34.36°C ± 3.17°C vs 68.6°C ± 3.7°C; P < .0001) and smaller thermal footprint (2.9 ± 0.2 mm vs 5.1 ± 0.2 mm; P < .0001), while maintaining perforation success compared with the "Cut" waveform. CONCLUSION: Displacement of the IAS is required for successful ETSP. Guidewire protrusion beyond 1 mm resulted in inconsistent perforation and increased tissue temperature.

Heart Rhythm
Royal Adelaide Hospital (AU), Flinders University (AU), The University of Adelaide (AU)
St. Jude Medical
Openalex Percentile: Top 0%
Thyroid and Parathyroid Surgery
26.17
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.