Evaluation of a Novel Multielectrode Surgical Pulsed Field Ablation Clamp: An Ex-Vivo Vegetal Study

OBJECTIVES: At present, there is no commercially available surgical device to perform cardiac ablation utilizing pulsed field ablation (PFA). We sought to explore the efficacy and requirements of a novel PFA clamp to achieve transmural lesions in a vegetal model of Tuberosum Solanum (TS). METHODS: We tested the PFA clamp at different voltages (1400 V, 2000 V, 2800 V) and numbers of applications (2 or 4) on slabs of TS having different thicknesses (10 mm, 20 mm, 30 mm). Each slab was sliced transversally and the halves put together or kept whole. Every possible combination was tested twice. After 72 hours, we observed the resulting necrotic tissue and calculated its volume. We assessed the determinants of %volume of necrotic tissue, defined as observed volume relative to the ideal volume of necrotic tissue. RESULTS: A transmural lesion was noted in 68 (94%) of 72 specimens. Using 4 applications, there was a strong linear relationship between voltage and %volume regardless of specimen thickness (R2 = 0.94, R2 = 0.96, R2 = 0.97, all P < .001) but not when using 2 applications (R2 = 0.37, R2 = 0.87, R2 = 0.69 for 10 mm, 20 mm, and 30 mm, respectively, all P < .05). The voltages at which 100% volume was observed using 4 applications-corresponding to the curve intersections-are 958 V, 928 V, and 1044 V for the 10 mm, 20 mm, and 30 mm specimens, respectively. CONCLUSIONS: The epicardial PFA clamp reliably produced transmural lesions in a vegetal model of TS using 2 or 4 applications. With 4 applications, lesion volume demonstrated a predictable, linear relationship with increasing voltage, supporting the device's potential for controlled and consistent ablation.

Authors

Institutions

Publication Details

Journal
Interdisciplinary CardioVascular and Thoracic Surgery
Published
2026-04-01
DOI
https://doi.org/10.1093/icvts/ivag089
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of a Novel Multielectrode Surgical Pulsed Field Ablation Clamp: An Ex-Vivo Vegetal Study

Alexis Puvrez, Ashutosh Hardikar, Melanie Gunawardene, Mohanaraj Jayakumar et al.
Interdisciplinary CardioVascular and Thoracic Surgery
Cardiac Arrhythmias and Treatments
article

Evaluation of a Novel Multielectrode Surgical Pulsed Field Ablation Clamp: An Ex-Vivo Vegetal Study

Alexis Puvrez, Ashutosh Hardikar, Melanie Gunawardene, Mohanaraj Jayakumar, Sanjai Pattu Valappil, Prashanthan Sanders, Ian Fong, J P Ariyaratnam, Mohamed Abbas, Jason Varzaly, Chris Robbins, Michael Kroon
article en

Abstract

OBJECTIVES: At present, there is no commercially available surgical device to perform cardiac ablation utilizing pulsed field ablation (PFA). We sought to explore the efficacy and requirements of a novel PFA clamp to achieve transmural lesions in a vegetal model of Tuberosum Solanum (TS). METHODS: We tested the PFA clamp at different voltages (1400 V, 2000 V, 2800 V) and numbers of applications (2 or 4) on slabs of TS having different thicknesses (10 mm, 20 mm, 30 mm). Each slab was sliced transversally and the halves put together or kept whole. Every possible combination was tested twice. After 72 hours, we observed the resulting necrotic tissue and calculated its volume. We assessed the determinants of %volume of necrotic tissue, defined as observed volume relative to the ideal volume of necrotic tissue. RESULTS: A transmural lesion was noted in 68 (94%) of 72 specimens. Using 4 applications, there was a strong linear relationship between voltage and %volume regardless of specimen thickness (R2 = 0.94, R2 = 0.96, R2 = 0.97, all P < .001) but not when using 2 applications (R2 = 0.37, R2 = 0.87, R2 = 0.69 for 10 mm, 20 mm, and 30 mm, respectively, all P < .05). The voltages at which 100% volume was observed using 4 applications-corresponding to the curve intersections-are 958 V, 928 V, and 1044 V for the 10 mm, 20 mm, and 30 mm specimens, respectively. CONCLUSIONS: The epicardial PFA clamp reliably produced transmural lesions in a vegetal model of TS using 2 or 4 applications. With 4 applications, lesion volume demonstrated a predictable, linear relationship with increasing voltage, supporting the device's potential for controlled and consistent ablation.

Interdisciplinary CardioVascular and Thoracic SurgeryVol. 41(4)
Royal Adelaide Hospital (AU), CathRx (Australia) (AU), The University of Adelaide (AU)
University of Adelaide, Medical Research Council, National Health and Medical Research Council, Centre for Heart Rhythm Disorders, University of Adelaide
Openalex Percentile: Top 26%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.