Feasibility of improving sLe(a)/CA 19-9 harmonization by recalibration: a simulation study using three EQA programs

OBJECTIVES: The CA 19-9 tumor marker measures the sialylated Lewis a blood group antigen (sLe(a)) expressed on circulating proteins. CA 19-9 is primarily used to monitor pancreatic and gastrointestinal cancers. Studies have demonstrated the lack of comparability of CA 19-9 assays. METHODS: Data from three EQA programs were used to simulate the potential for increasing the CA 19-9 harmonization. The simulation was based on method comparison studies of each pair of measurement procedures (MPs), selection of comparable MPs for value assignment of EQA samples, and mathematical recalibration of MPs. Harmonization status was determined pre- and post-harmonization, using consensus mean of all MPs, as reference, and method comparison studies. RESULTS: Seventy-five EQA samples were included and four MPs (Abbott, Beckman, Roche and Siemens (Atellica)) were present in all EQA programs and five others from Siemens (Immulite), Ortho, Fujirebio, Biomerieux and Thermo Fisher, in at least one EQA program. The obtained harmonization procedure required recalibration of all MPs. Post-harmonization mean difference improved from -50.3 to 92.0 % into -0.48-1.25 %. The range of slopes from individual MP method comparisons improved from 0.090 to 2.56 into 0.93 to 1.10. Post-harmonization CV(%) of individual EQA samples remained at 22.0-49.2 % within individual method comparisons and alignment of upper limits of normal diversified after application of the harmonization procedure. CONCLUSIONS: Overall, CA 19-9 harmonization can be improved by recalibration, though a significant lack of correlation remains. The feasibility of achieving CA 19-9 harmonization of today's clinical assays is questionable.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-04-14
DOI
https://doi.org/10.1515/cclm-2026-0155
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of improving sLe(a)/CA 19-9 harmonization by recalibration: a simulation study using three EQA programs

Daan van de Kerkhof, Huub H. van Rossum, Yeo‐Min Yun, Catharine M. Sturgeon et al.
Clinical Chemistry and Laboratory Medicine (CCLM)
Cardiac Imaging and Diagnostics
article

Feasibility of improving sLe(a)/CA 19-9 harmonization by recalibration: a simulation study using three EQA programs

Daan van de Kerkhof, Huub H. van Rossum, Yeo‐Min Yun, Catharine M. Sturgeon, Vathany Kulasingam, Xu Qian, Hubert W. Vesper, Stefan Holdenrieder, Qing H. Meng, Marc Thelen, Wei Cui, Katherine Partridge, Dina Patel, Adrianna Herskovits, Melanie Moore, Junghan Song, Bart E.P.B. Ballieux
article en

Abstract

OBJECTIVES: The CA 19-9 tumor marker measures the sialylated Lewis a blood group antigen (sLe(a)) expressed on circulating proteins. CA 19-9 is primarily used to monitor pancreatic and gastrointestinal cancers. Studies have demonstrated the lack of comparability of CA 19-9 assays. METHODS: Data from three EQA programs were used to simulate the potential for increasing the CA 19-9 harmonization. The simulation was based on method comparison studies of each pair of measurement procedures (MPs), selection of comparable MPs for value assignment of EQA samples, and mathematical recalibration of MPs. Harmonization status was determined pre- and post-harmonization, using consensus mean of all MPs, as reference, and method comparison studies. RESULTS: Seventy-five EQA samples were included and four MPs (Abbott, Beckman, Roche and Siemens (Atellica)) were present in all EQA programs and five others from Siemens (Immulite), Ortho, Fujirebio, Biomerieux and Thermo Fisher, in at least one EQA program. The obtained harmonization procedure required recalibration of all MPs. Post-harmonization mean difference improved from -50.3 to 92.0 % into -0.48-1.25 %. The range of slopes from individual MP method comparisons improved from 0.090 to 2.56 into 0.93 to 1.10. Post-harmonization CV(%) of individual EQA samples remained at 22.0-49.2 % within individual method comparisons and alignment of upper limits of normal diversified after application of the harmonization procedure. CONCLUSIONS: Overall, CA 19-9 harmonization can be improved by recalibration, though a significant lack of correlation remains. The feasibility of achieving CA 19-9 harmonization of today's clinical assays is questionable.

Clinical Chemistry and Laboratory Medicine (CCLM)
Centers for Disease Control and Prevention (US), Memorial Sloan Kettering Cancer Center (US), The University of Texas MD Anderson Cancer Center (US), Radboud University Nijmegen (NL), Environment Agency (GB), University of Toronto (CA), Chinese Academy of Medical Sciences & Peking Union Medical College (CN), Leiden University Medical Center (NL), Catharina Ziekenhuis (NL), Northern General Hospital (GB), Edinburgh Royal Infirmary (GB), Radboud University Medical Center (NL), Seoul National University Bundang Hospital (KR), The Netherlands Cancer Institute (NL), Konkuk University Medical Center (KR), National Center for Environmental Health (US), Zhejiang Cancer Hospital (CN), Streekziekenhuis Koningin Beatrix (NL), Deutsches Herzzentrum München (DE)
Openalex Percentile: Top 20%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.