NATON robotic-assisted unicompartmental knee arthroplasty achieves superior radiological alignment and comparable one-year functional outcomes to conventional surgery

To compare the early clinical efficacy and postoperative implant alignment between NATON robotic-assisted unicompartmental knee arthroplasty (RA-UKA) and conventional UKA (C-UKA) in treating severe medial compartment knee osteoarthritis (OA). This retrospective matched cohort study included 40 RA-UKA and 40 C-UKA patients, matched 1:1 by age, gender, and surgical site et al. Primary outcomes included operative time, intraoperative blood loss, and patient-reported outcome measures (PROMs): knee society score (KSS), Oxford knee score (OKS), visual analogue scale (VAS) for pain, and forgotten joint score (FJS). Radiological assessments included coronal/sagittal angular deviations of femoral/tibial components, hip-knee-ankle (HKA) angle, and Kennedy zone distribution. The follow-up time was 15.4 ± 1.6 months. Operative time was significantly longer in the RA-UKA group (95.8 ± 6.8 min vs. 75.3 ± 6.5 min, P < 0.001). No significant differences were observed in KSS score, OKS score, or intraoperative blood loss between groups (all P > 0.05). RA-UKA achieved significantly better subjective PROMs, with lower VAS score (1.97 vs. 2.5, P = 0.012, Cohen’s d = − 0.60) and higher FJS score (86.9 vs. 84.3, P = 0.025, Cohen’s d = 0.50). The RA-UKA group had smaller femoral component coronal alignment deviation (2.39 ± 1.41° vs. 5.79 ± 1.15°, P < 0.001, Cohen’s d = − 2.46) and flexion alignment deviation (3.3° (2.9–4.0) vs. 6.5° (5.9–7.1), P < 0.001, Cohen’s d = − 0.78). No significant differences were observed in tibial component coronal deviation and posterior tibial slope(PTS) deviation between two groups (all P > 0.05). Correlation analysis revealed weak negative correlations between FJS score and femoral coronal/flexion deviations (all P < 0.05). VAS score was weakly positive correlated with PTS deviation (ρ = 0.240, P = 0.032) and moderately positive correlated with femoral coronal deviation (r = 0.311, P = 0.005). No serious complications occurred in either group during follow-up. The NATON RA-UKA achieved more accurate early implant positioning than C-UKA, while short-term clinical outcomes were generally comparable between groups.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-04-14
DOI
https://doi.org/10.1038/s41598-026-48561-9
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

NATON robotic-assisted unicompartmental knee arthroplasty achieves superior radiological alignment and comparable one-year functional outcomes to conventional surgery

尹毅, 吴菊, Guanjun Sun, Ke He et al.
Scientific Reports
Total Knee Arthroplasty Outcomes
article

NATON robotic-assisted unicompartmental knee arthroplasty achieves superior radiological alignment and comparable one-year functional outcomes to conventional surgery

尹毅, 吴菊, Guanjun Sun, Ke He, Xu Peng, Yuan Li
article en

Abstract

To compare the early clinical efficacy and postoperative implant alignment between NATON robotic-assisted unicompartmental knee arthroplasty (RA-UKA) and conventional UKA (C-UKA) in treating severe medial compartment knee osteoarthritis (OA). This retrospective matched cohort study included 40 RA-UKA and 40 C-UKA patients, matched 1:1 by age, gender, and surgical site et al. Primary outcomes included operative time, intraoperative blood loss, and patient-reported outcome measures (PROMs): knee society score (KSS), Oxford knee score (OKS), visual analogue scale (VAS) for pain, and forgotten joint score (FJS). Radiological assessments included coronal/sagittal angular deviations of femoral/tibial components, hip-knee-ankle (HKA) angle, and Kennedy zone distribution. The follow-up time was 15.4 ± 1.6 months. Operative time was significantly longer in the RA-UKA group (95.8 ± 6.8 min vs. 75.3 ± 6.5 min, P < 0.001). No significant differences were observed in KSS score, OKS score, or intraoperative blood loss between groups (all P > 0.05). RA-UKA achieved significantly better subjective PROMs, with lower VAS score (1.97 vs. 2.5, P = 0.012, Cohen’s d = − 0.60) and higher FJS score (86.9 vs. 84.3, P = 0.025, Cohen’s d = 0.50). The RA-UKA group had smaller femoral component coronal alignment deviation (2.39 ± 1.41° vs. 5.79 ± 1.15°, P < 0.001, Cohen’s d = − 2.46) and flexion alignment deviation (3.3° (2.9–4.0) vs. 6.5° (5.9–7.1), P < 0.001, Cohen’s d = − 0.78). No significant differences were observed in tibial component coronal deviation and posterior tibial slope(PTS) deviation between two groups (all P > 0.05). Correlation analysis revealed weak negative correlations between FJS score and femoral coronal/flexion deviations (all P < 0.05). VAS score was weakly positive correlated with PTS deviation (ρ = 0.240, P = 0.032) and moderately positive correlated with femoral coronal deviation (r = 0.311, P = 0.005). No serious complications occurred in either group during follow-up. The NATON RA-UKA achieved more accurate early implant positioning than C-UKA, while short-term clinical outcomes were generally comparable between groups.

Scientific Reports
Suizhou Central Hospital (CN)
Good health and well-being
Openalex Percentile: Top 9%
Total Knee Arthroplasty Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.