Sensitivity and concordance study of an open source EPID‐based linear accelerator test suite

BACKGROUND: Electronic portal imaging devices (EPIDs) enhance linear accelerator (linac) quality assurance (QA) efficiency but are not universally used beyond patient-specific IMRT QA. A multi-institutional Consortium seeks to standardize linac QA utilizing an EPID-based test suite. PURPOSE: This study evaluates the sensitivity of a test suite to intentional linac and plan errors and the concordance with linac fluctuations over time. METHODS: Baseline and error-introduced tests were performed on a Varian TrueBeam and Clinac. Evaluated parameters included enhanced dynamic wedge factors (EDW) (6MV), central axis dose (6MV, 6MV FFF, 16MV), focal spot alignment and beam symmetry (6MV, 16MV), dose rate-gantry speed and multi-leaf collimator (MLC) leaf speed (6MV). Errors included varying EDW angles (10°-60°), output changes, beam steering for focal spot and symmetry inaccuracies, and meterset adjustments of control points for dose rate-gantry speed and MLC leaf speed tests. Measurements were performed with an EPID, analyzed via test suite software, and compared to an ion chamber (IC) or array. Concordance analysis was performed on a single linac over 5 months to assess the test suite's ability to quantify deviations during routine operations. Measurement reproducibility was assessed. RESULTS: Results were reported as differences between EPID and an IC or array. For each EDW, wedge factor differences were <0.5% (TrueBeam) and <1.5% (Clinac). Central axis dose variations stayed within ±0.8%. For focal spot, EPID underestimated the radial direction and overestimated the transverse direction. Symmetry measurements showed strong linearity (r > 0.99), though EPID measurements underestimated changes in symmetry relative to baseline. Maximum differences for dose rate-gantry speed and MLC leaf speed were <0.4%. In the concordance study, output differences averaged 0.3% ± 0.3% (6MV), 0.5% ± 0.3% (6MV FFF), and 0.3% ± 0.2% (16MV). Systematic differences in symmetry were observed between EPID and a 2D-array with standard deviations <0.2%. CONCLUSIONS: Comparisons with traditional detectors showed the EPID test suite can detect errors and supports commissioning and clinical integration.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-04-01
DOI
https://doi.org/10.1002/acm2.70569
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sensitivity and concordance study of an open source EPID‐based linear accelerator test suite

Seng Boh Lim, Lana C. Critchfield, Cory Knill, Richard A. Popple et al.
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Sensitivity and concordance study of an open source EPID‐based linear accelerator test suite

Seng Boh Lim, Lana C. Critchfield, Cory Knill, Richard A. Popple, Michael P. Barnes, Jean Moran, William Donahue, Nilendu Gupta, John J. DeMarco, Ellie Durussel, Jim Irrer, Mario Perez, Hania Al‐Hallaq, Justin K. Mikell, Michael Pillainayagam
article en

Abstract

BACKGROUND: Electronic portal imaging devices (EPIDs) enhance linear accelerator (linac) quality assurance (QA) efficiency but are not universally used beyond patient-specific IMRT QA. A multi-institutional Consortium seeks to standardize linac QA utilizing an EPID-based test suite. PURPOSE: This study evaluates the sensitivity of a test suite to intentional linac and plan errors and the concordance with linac fluctuations over time. METHODS: Baseline and error-introduced tests were performed on a Varian TrueBeam and Clinac. Evaluated parameters included enhanced dynamic wedge factors (EDW) (6MV), central axis dose (6MV, 6MV FFF, 16MV), focal spot alignment and beam symmetry (6MV, 16MV), dose rate-gantry speed and multi-leaf collimator (MLC) leaf speed (6MV). Errors included varying EDW angles (10°-60°), output changes, beam steering for focal spot and symmetry inaccuracies, and meterset adjustments of control points for dose rate-gantry speed and MLC leaf speed tests. Measurements were performed with an EPID, analyzed via test suite software, and compared to an ion chamber (IC) or array. Concordance analysis was performed on a single linac over 5 months to assess the test suite's ability to quantify deviations during routine operations. Measurement reproducibility was assessed. RESULTS: Results were reported as differences between EPID and an IC or array. For each EDW, wedge factor differences were <0.5% (TrueBeam) and <1.5% (Clinac). Central axis dose variations stayed within ±0.8%. For focal spot, EPID underestimated the radial direction and overestimated the transverse direction. Symmetry measurements showed strong linearity (r > 0.99), though EPID measurements underestimated changes in symmetry relative to baseline. Maximum differences for dose rate-gantry speed and MLC leaf speed were <0.4%. In the concordance study, output differences averaged 0.3% ± 0.3% (6MV), 0.5% ± 0.3% (6MV FFF), and 0.3% ± 0.2% (16MV). Systematic differences in symmetry were observed between EPID and a 2D-array with standard deviations <0.2%. CONCLUSIONS: Comparisons with traditional detectors showed the EPID test suite can detect errors and supports commissioning and clinical integration.

Journal of Applied Clinical Medical PhysicsVol. 27(4)
University of Essex (GB), Cedars-Sinai Medical Center (US), Memorial Sloan Kettering Cancer Center (US), Emory University (US), Washington University in St. Louis (US), University of Michigan (US), The Ohio State University Comprehensive Cancer Center – Arthur G. James Cancer Hospital and Richard J. Solove Research Institute (US), Metro Health Hospital (US), University of Alabama at Birmingham (US), Northern Sydney Local Health District (AU), Calvary Mater Newcastle Hospital (AU)
Varian Medical Systems
Openalex Percentile: Top 16%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.