A novel urine collection device for accurate urinary measurements in neonates and pre-continent infants

BACKGROUND: Urine collection from neonates and infants is challenging. All available urine collection methods (UCMs) have their disadvantages, like invasiveness, contamination, and/or inaccurate measurements of urinary components. This study aimed to demonstrate the accuracy of measurements of urinary components for the novel non-invasive 'Specturi urine collector device'. METHODS: The urine collection textile was incubated under different conditions (incubation duration, temperature, and humidity) with either artificial or human urine. Concentrations of urinary components were compared before and after incubation. Because of suspected crystal formation, the effect of acidification was assessed in combination with crystal analysis. RESULTS: Urinary concentrations of calcium, magnesium, phosphate, sodium, creatinine, and protein, and pH were accurately measured, except for the calcium concentration in alkaline urines. The incubation conditions showed no to a small effect on the measurement of urinary components, which does not affect clinical decision-making. Crystals containing calcium, magnesium, and phosphorus formed on the textile after incubation with alkaline urine. Acidification recovered the calcium concentration in these urines. CONCLUSION: The non-invasive 'Specturi urine collector device' provides accurate measurements of urinary components under different conditions. This UCM is more accurate than other textile-based UCMs. Eventually, this device can be used as an accurate as well as patient- and staff-friendly UCM. IMPACT: All urine collection methods for neonates and pre-continent infants have their disadvantages, and textile-based urine collection methods provide inaccurate measurements of urinary components. Furthermore, the effect of changing environmental conditions on measurements of urinary components is unknown for these urine collection methods. This study introduces the 'Specturi urine collector device', a novel non-invasive urine collection device. This device provides accurate measurements for different urinary components under different conditions, like incubation duration, temperature, and humidity. The results show that the 'Specturi urine collector device' has the potential to become an alternative non-invasive urine collection method that is accurate, and patient- and staff-friendly.

Authors

Institutions

Publication Details

Journal
Pediatric Research
Published
2026-04-13
DOI
https://doi.org/10.1038/s41390-026-04891-y
Primary Topic
Urinary Tract Infections Management
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel urine collection device for accurate urinary measurements in neonates and pre-continent infants

Miranda van Berkel, Joost G.J. Hoenderop, S. H. M. Nauta, Christine ten Hove et al.
Pediatric Research
Urinary Tract Infections Management
article

A novel urine collection device for accurate urinary measurements in neonates and pre-continent infants

Miranda van Berkel, Joost G.J. Hoenderop, S. H. M. Nauta, Christine ten Hove, Petra van Setten, Lieke A.E.A. Corbeek
article en

Abstract

BACKGROUND: Urine collection from neonates and infants is challenging. All available urine collection methods (UCMs) have their disadvantages, like invasiveness, contamination, and/or inaccurate measurements of urinary components. This study aimed to demonstrate the accuracy of measurements of urinary components for the novel non-invasive 'Specturi urine collector device'. METHODS: The urine collection textile was incubated under different conditions (incubation duration, temperature, and humidity) with either artificial or human urine. Concentrations of urinary components were compared before and after incubation. Because of suspected crystal formation, the effect of acidification was assessed in combination with crystal analysis. RESULTS: Urinary concentrations of calcium, magnesium, phosphate, sodium, creatinine, and protein, and pH were accurately measured, except for the calcium concentration in alkaline urines. The incubation conditions showed no to a small effect on the measurement of urinary components, which does not affect clinical decision-making. Crystals containing calcium, magnesium, and phosphorus formed on the textile after incubation with alkaline urine. Acidification recovered the calcium concentration in these urines. CONCLUSION: The non-invasive 'Specturi urine collector device' provides accurate measurements of urinary components under different conditions. This UCM is more accurate than other textile-based UCMs. Eventually, this device can be used as an accurate as well as patient- and staff-friendly UCM. IMPACT: All urine collection methods for neonates and pre-continent infants have their disadvantages, and textile-based urine collection methods provide inaccurate measurements of urinary components. Furthermore, the effect of changing environmental conditions on measurements of urinary components is unknown for these urine collection methods. This study introduces the 'Specturi urine collector device', a novel non-invasive urine collection device. This device provides accurate measurements for different urinary components under different conditions, like incubation duration, temperature, and humidity. The results show that the 'Specturi urine collector device' has the potential to become an alternative non-invasive urine collection method that is accurate, and patient- and staff-friendly.

Pediatric Research
Radboud University Nijmegen (NL), Radboud University Medical Center (NL), Radboud Institute for Molecular Life Sciences (NL), Amalia Kinderziekenhuis (NL), Rijnstate Hospital (NL)
Health~Holland, Radboud Universiteit, Nierstichting, Radboud Universitair Medisch Centrum
Openalex Percentile: Top 17%
Urinary Tract Infections Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.