Dosimetric Performance Analysis of Pediatric Based Halcyon™ Volumetric Arc Therapy-Based Lung-Sparing Total Body Irradiation

IntroductionHalcyon™ O-ring linear accelerators (linacs) provide a robust and efficient radiotherapy option for clinics. Interest in volumetric arc therapy (VMAT) based total body irradiation (TBI) has grown, with most current VMAT TBI protocols employing C-arm linacs; however, the feasibility of Halcyon-based VMAT TBI has been recently hypothesized. This study investigates the planning feasibility and dosimetric quality of VMAT-based, lung-sparing TBI delivered by Halcyon linac to pediatric patients.Methods10 pediatric patients (94 - 134 cm in height) previously treated to 12 Gy TBI through traditional 3D planning techniques were retrospectively replanned on a Halcyon linac. 2-3 VMAT plans containing 4-6 isocenters each with 2-2.5 arcs were created to treat the head, abdomen, and legs, with 1-2 static-gantry treatment matching areas being used between VMAT planning regions.ResultsLung doses were significantly reduced below the clinically significant level of 8 Gy, achieving mean doses of 6.66 ± 0.52 Gy (55.5% ± 4.3%). Target coverage was achieved with D90% of 77.5% ± 12.5% and homogeneity indexes of 0.92 ± 0.1. Planning time took 5-6 hours per patient: 1.5 hours for contouring and field placement and 3.5-4.5 hours for plan optimization and dose calculations. Using a dose rate of 800 MU/min field delivery time took an average of 20.8 ± 7 minutes with average portal dosimetry quality assurance passing rates of 99.6% and 99.9% for 2%/2mm and 3%/3mm gamma analysis metrics.ConclusionVMAT lung sparing TBI via Halcyon linacs is feasible and dosimetrically acceptable, with greatest applicability to smaller patients (<100 cm). Planning and setup complexity increases dramatically for larger patients (>100 cm) due to the smaller maximum field sizes (28 x 28 cm2) achievable. Future developments in software-based automations for target delineation, multiple field matching, and plan optimization could allow for more practicable treatment of larger patients.

Authors

Institutions

Publication Details

Journal
Technology in Cancer Research & Treatment
Published
2026-04-01
DOI
https://doi.org/10.1177/15330338261443068
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dosimetric Performance Analysis of Pediatric Based Halcyon™ Volumetric Arc Therapy-Based Lung-Sparing Total Body Irradiation

Christine E. Hill‐Kayser, Allison Haertter, Michael J. LaRiviere, Michele M. Kim et al.
Technology in Cancer Research & Treatment
Advanced Radiotherapy Techniques
article

Dosimetric Performance Analysis of Pediatric Based Halcyon™ Volumetric Arc Therapy-Based Lung-Sparing Total Body Irradiation

Christine E. Hill‐Kayser, Allison Haertter, Michael J. LaRiviere, Michele M. Kim, John P. Plastaras, Jacob Niculcea, Taoran Li, Keshav Goel
article en

Abstract

IntroductionHalcyon™ O-ring linear accelerators (linacs) provide a robust and efficient radiotherapy option for clinics. Interest in volumetric arc therapy (VMAT) based total body irradiation (TBI) has grown, with most current VMAT TBI protocols employing C-arm linacs; however, the feasibility of Halcyon-based VMAT TBI has been recently hypothesized. This study investigates the planning feasibility and dosimetric quality of VMAT-based, lung-sparing TBI delivered by Halcyon linac to pediatric patients.Methods10 pediatric patients (94 - 134 cm in height) previously treated to 12 Gy TBI through traditional 3D planning techniques were retrospectively replanned on a Halcyon linac. 2-3 VMAT plans containing 4-6 isocenters each with 2-2.5 arcs were created to treat the head, abdomen, and legs, with 1-2 static-gantry treatment matching areas being used between VMAT planning regions.ResultsLung doses were significantly reduced below the clinically significant level of 8 Gy, achieving mean doses of 6.66 ± 0.52 Gy (55.5% ± 4.3%). Target coverage was achieved with D90% of 77.5% ± 12.5% and homogeneity indexes of 0.92 ± 0.1. Planning time took 5-6 hours per patient: 1.5 hours for contouring and field placement and 3.5-4.5 hours for plan optimization and dose calculations. Using a dose rate of 800 MU/min field delivery time took an average of 20.8 ± 7 minutes with average portal dosimetry quality assurance passing rates of 99.6% and 99.9% for 2%/2mm and 3%/3mm gamma analysis metrics.ConclusionVMAT lung sparing TBI via Halcyon linacs is feasible and dosimetrically acceptable, with greatest applicability to smaller patients (<100 cm). Planning and setup complexity increases dramatically for larger patients (>100 cm) due to the smaller maximum field sizes (28 x 28 cm2) achievable. Future developments in software-based automations for target delineation, multiple field matching, and plan optimization could allow for more practicable treatment of larger patients.

Technology in Cancer Research & TreatmentVol. 25
New York Proton Center (US), University of Pennsylvania (US)
GE Healthcare
Openalex Percentile: Top 21%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.