Improving patient treatment accuracy using transit dosimetry with Electronic Portal Imaging Device images and deep learning
Background and Purpose: Electronic portal imaging devices (EPID) can be utilized for transit in vivo dosimetry (IVD). A major limitation of standard amorphous Silicon EPID detectors for usage IVD is their non-linear detector response to water equivalent dose. Labor-intensive calibration and correction procedures are needed, limiting their practicality for routine clinical use. The purpose of this work was to develop a deep learning (DL) framework, based on data acquisitions and treatment planning system (TPS) dose simulations, which reconstructed water equivalent dose from transit EPID images. Materials and methods: -index analysis with a 3%/3 mm and 5%/5 mm criterion. Results: for the 3%/3 mm and 5%/5 mm criteria, respectively. Median passing rates were 98.3% for the 3%/3 mm and 100% for the 5%/5 mm criterion, respectively, demonstrating high accuracy in reconstructing portal dose from EPID. The predictions were completed within 1 s. Conclusions: Our DL framework enabled fast and accurate reconstruction of portal dose from transit EPID images. Once trained on clinical data, it may reduce complex corrections, supporting safe and effective treatments that meet modern regulatory standards.
Authors
- G. Pirrone (ORCID: https://orcid.org/0000-0002-6113-6644)
- Michele Avanzo (ORCID: https://orcid.org/0000-0003-1711-4242)
- Emmanuel Uwitonze
- Francesca Lizzi (ORCID: https://orcid.org/0000-0003-0900-0421)
- Alessandra Retico (ORCID: https://orcid.org/0000-0001-5135-4472)
- Lorenzo Marini
- A. Kraan (ORCID: https://orcid.org/0000-0002-1417-8809)
- Cinzia Talamonti
- Icro Meattini
- Carlotta Mozzi
Institutions
- University of Pisa (IT)
- Istituto Nazionale di Fisica Nucleare (IT)
- Maastricht University Medical Centre (NL)
- Azienda Ospedaliero-Universitaria Careggi (IT)
- Centro di Riferimento Oncologico (IT)
- Rwanda Biomedical Center (RW)
- Istituti di Ricovero e Cura a Carattere Scientifico (IT)
- Istituto Nazionale di Fisica Nucleare, Sezione di Pisa (IT)
- Maastro Clinic (NL)
- University of Florence (IT)
Publication Details
- Journal
- Physics and Imaging in Radiation Oncology
- Published
- 2026-04-13
- DOI
- https://doi.org/10.1016/j.phro.2026.100966
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Instituto Nazionale di Fisica Nucleare