Retrieval of a Free-Floating Transcatheter Edge-to-Edge Repair (TEER) Clip in a Giant Left Atrium

BACKGROUND: Device embolization during transcatheter edge-to-edge repair is a rare and catastrophic complication. Retrieval can be particularly challenging in a giant left atrium (LA), where jet-driven device motion may render active "chasing" ineffective. CASE SUMMARY: A DragonFly clip completely detached in a 75-year-old man with a giant LA (anteroposterior diameter approximately 10 cm). The device moved erratically driven by eccentric regurgitant jets, making active pursuit unsuccessful. The operators switched to a fluoroscopy-guided passive capture strategy after identifying a consistent turning/impact zone on the lateral atrial wall. The snare was stationed at this specific site to await and intercept the recirculating clip, which allowed for subsequent orientation and atraumatic extraction. DISCUSSION: This case illustrates a practical "floating clip" phenomenon in a giant LA. When fluoroscopy demonstrates a repeatable device trajectory, passive capture at a reproducible turning/impact zone may facilitate device retrieval and support continued percutaneous management in selected cases.

Authors

Institutions

Publication Details

Journal
JACC Case Reports
Published
2026-04-01
DOI
https://doi.org/10.1016/j.jaccas.2026.108026
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Retrieval of a Free-Floating Transcatheter Edge-to-Edge Repair (TEER) Clip in a Giant Left Atrium

Zhongshan Gou, Iokfai Cheang, Tianbao Xu, Haifeng Zhang et al.
JACC Case Reports
Cardiac Arrest and Resuscitation
article

Retrieval of a Free-Floating Transcatheter Edge-to-Edge Repair (TEER) Clip in a Giant Left Atrium

Zhongshan Gou, Iokfai Cheang, Tianbao Xu, Haifeng Zhang, Ruitao Hou, Jiayi Weng, Munire Zaibi
article en

Abstract

BACKGROUND: Device embolization during transcatheter edge-to-edge repair is a rare and catastrophic complication. Retrieval can be particularly challenging in a giant left atrium (LA), where jet-driven device motion may render active "chasing" ineffective. CASE SUMMARY: A DragonFly clip completely detached in a 75-year-old man with a giant LA (anteroposterior diameter approximately 10 cm). The device moved erratically driven by eccentric regurgitant jets, making active pursuit unsuccessful. The operators switched to a fluoroscopy-guided passive capture strategy after identifying a consistent turning/impact zone on the lateral atrial wall. The snare was stationed at this specific site to await and intercept the recirculating clip, which allowed for subsequent orientation and atraumatic extraction. DISCUSSION: This case illustrates a practical "floating clip" phenomenon in a giant LA. When fluoroscopy demonstrates a repeatable device trajectory, passive capture at a reproducible turning/impact zone may facilitate device retrieval and support continued percutaneous management in selected cases.

JACC Case Reports
Xinjiang Medical University (CN), Suzhou Municipal Hospital (CN), Nanjing Medical University (CN)
Openalex Percentile: Top 22%
Cardiac Arrest and Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.