Retrieval of a Free-Floating Transcatheter Edge-to-Edge Repair (TEER) Clip in a Giant Left Atrium
BACKGROUND: Device embolization during transcatheter edge-to-edge repair is a rare and catastrophic complication. Retrieval can be particularly challenging in a giant left atrium (LA), where jet-driven device motion may render active "chasing" ineffective. CASE SUMMARY: A DragonFly clip completely detached in a 75-year-old man with a giant LA (anteroposterior diameter approximately 10 cm). The device moved erratically driven by eccentric regurgitant jets, making active pursuit unsuccessful. The operators switched to a fluoroscopy-guided passive capture strategy after identifying a consistent turning/impact zone on the lateral atrial wall. The snare was stationed at this specific site to await and intercept the recirculating clip, which allowed for subsequent orientation and atraumatic extraction. DISCUSSION: This case illustrates a practical "floating clip" phenomenon in a giant LA. When fluoroscopy demonstrates a repeatable device trajectory, passive capture at a reproducible turning/impact zone may facilitate device retrieval and support continued percutaneous management in selected cases.
Authors
- Zhongshan Gou (ORCID: https://orcid.org/0000-0003-4403-0765)
- Iokfai Cheang (ORCID: https://orcid.org/0000-0002-9950-9696)
- Tianbao Xu
- Haifeng Zhang
- Ruitao Hou (ORCID: https://orcid.org/0000-0001-9226-7535)
- Jiayi Weng
- Munire Zaibi
Institutions
- Xinjiang Medical University (CN)
- Suzhou Municipal Hospital (CN)
- Nanjing Medical University (CN)
Publication Details
- Journal
- JACC Case Reports
- Published
- 2026-04-01
- DOI
- https://doi.org/10.1016/j.jaccas.2026.108026
- Primary Topic
- Cardiac Arrest and Resuscitation
- Type
- article
- Field-Weighted Citation Impact
- 0.00