Continuous bed motion versus step-and-shoot acquisition in LAFOV PET/CT: insights from multi-phantom and patient studies

BACKGROUND: Continuous bed motion (CBM) allows flexible extension of the scan range compared to conventional step‑and‑shoot (S&S) acquisition but has not yet been evaluated in long axial field‑of‑view (LAFOV) PET/CT. This study systematically assessed the impact of CBM on image quality, noise, and quantitative performance in the Biograph Vision Quadra LAFOV PET/CT using multi‑phantom and patient scans compared to S&S. METHODS: F]FDG underwent sequential CBM (2.8 mm/s, 378 s) and S&S (300 s) scans. Image noise, net true counts, and liver and lesion SUV values were compared using paired statistics and Bland-Altman analysis, along with PSMA expression scores. RESULTS: bias - 0.1 (liver) and - 0.8 (lesions)) did not affect clinical scores. CONCLUSIONS: The prolonged CBM protocol provides image quality and quantitative performance comparable to S&S in LAFOV PET/CT. While the reconstructed image range remains constrained by CT coverage, the patient scan comparison with 106 cm scan range, together with extended range phantom measurements, indicate that CBM can support scan range extension beyond 106 cm without compromising diagnostic accuracy.

Authors

Institutions

Publication Details

Journal
EJNMMI Physics
Published
2026-04-12
DOI
https://doi.org/10.1186/s40658-026-00867-3
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Continuous bed motion versus step-and-shoot acquisition in LAFOV PET/CT: insights from multi-phantom and patient studies

Christian la Fougère, Eduardo Calderón, Ivo Rausch, Pia M. Linder et al.
EJNMMI Physics
Medical Imaging Techniques and Applications
article

Continuous bed motion versus step-and-shoot acquisition in LAFOV PET/CT: insights from multi-phantom and patient studies

Christian la Fougère, Eduardo Calderón, Ivo Rausch, Pia M. Linder, Fabian Schmidt, Wenhong Lan
article en

Abstract

BACKGROUND: Continuous bed motion (CBM) allows flexible extension of the scan range compared to conventional step‑and‑shoot (S&S) acquisition but has not yet been evaluated in long axial field‑of‑view (LAFOV) PET/CT. This study systematically assessed the impact of CBM on image quality, noise, and quantitative performance in the Biograph Vision Quadra LAFOV PET/CT using multi‑phantom and patient scans compared to S&S. METHODS: F]FDG underwent sequential CBM (2.8 mm/s, 378 s) and S&S (300 s) scans. Image noise, net true counts, and liver and lesion SUV values were compared using paired statistics and Bland-Altman analysis, along with PSMA expression scores. RESULTS: bias - 0.1 (liver) and - 0.8 (lesions)) did not affect clinical scores. CONCLUSIONS: The prolonged CBM protocol provides image quality and quantitative performance comparable to S&S in LAFOV PET/CT. While the reconstructed image range remains constrained by CT coverage, the patient scan comparison with 106 cm scan range, together with extended range phantom measurements, indicate that CBM can support scan range extension beyond 106 cm without compromising diagnostic accuracy.

EJNMMI PhysicsVol. 13(1)
Medical University of Vienna (AT), University of Tübingen (DE)
Siemens Medical Solutions USA, Universitätsklinikum Tübingen
Industry, innovation and infrastructure
Openalex Percentile: Top 22%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.