Development and biomechanical validation of a 3D-printed Schepens scleral depressor

This study aimed to develop a low-cost 3D-printed scleral depressor and evaluate its mechanical performance, safety margins, and ocular biomechanical effects. A Schepens-style depressor was developed and printed in PLA. Examiners performed two different tests: (1) the maximum simulated scleral depression force, using both the 3D-printed and conventional steel depressors, and (2) a breakage test performed only on the 3D-printed device, determining its mechanical failure threshold for probabilistic safety analysis. Peak forces were applied to the porcine belly and recorded by a precision balance with slow-motion video analysis. A third test, which was conducted exclusively with the 3D-printed depressor, was performed using one ex vivo porcine eye model to correlate the applied force with the induced intraocular pressure (IOP) elevation. The pressure-volume behavior was modeled via the Friedenwald rigidity coefficient. One unit of the depressor prototype consumed 3.06 g of PLA, with an estimated cost and print time of U$ 0.06 and 22 min, respectively. The simulated indentations produced forces of 21.21 ± 6.23 N (3D-printed depressor) and 25.02 ± 4.64 N (steel depressor), with no significant difference between devices. The 3D-printed instrument breakage point was 63.27 ± 10.72 N, with a 2.98 factor of safety (FS) and 3.39 reliability index (β). In the porcine model, scleral depression produced a 15.63 ± 8.13 mmHg increase in IOP, requiring 0.191 ± 0.09 N (FS = 331.2 and β = 5.88). The 3D-printed depressor demonstrates effective mechanical robustness, wide safety margins, and functional equivalence to steel instruments, supporting the use of customizable, low-cost 3D-printed depressors in training and clinical settings.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2026-04-11
DOI
https://doi.org/10.1177/09544119261441042
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and biomechanical validation of a 3D-printed Schepens scleral depressor

João Pedro Vieira Neto Murta, Luís Expedito Sabage, Josmar Sabage, G. Manfredini et al.
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Glaucoma and retinal disorders
article

Development and biomechanical validation of a 3D-printed Schepens scleral depressor

João Pedro Vieira Neto Murta, Luís Expedito Sabage, Josmar Sabage, G. Manfredini, Alessandra Mazzo, Mariana Pasqualin Wojcikiewicz, Carlos Augusto Moreira-Neto, Jair Marcelo Saad Ortega, Thiago Meister
article en

Abstract

This study aimed to develop a low-cost 3D-printed scleral depressor and evaluate its mechanical performance, safety margins, and ocular biomechanical effects. A Schepens-style depressor was developed and printed in PLA. Examiners performed two different tests: (1) the maximum simulated scleral depression force, using both the 3D-printed and conventional steel depressors, and (2) a breakage test performed only on the 3D-printed device, determining its mechanical failure threshold for probabilistic safety analysis. Peak forces were applied to the porcine belly and recorded by a precision balance with slow-motion video analysis. A third test, which was conducted exclusively with the 3D-printed depressor, was performed using one ex vivo porcine eye model to correlate the applied force with the induced intraocular pressure (IOP) elevation. The pressure-volume behavior was modeled via the Friedenwald rigidity coefficient. One unit of the depressor prototype consumed 3.06 g of PLA, with an estimated cost and print time of U$ 0.06 and 22 min, respectively. The simulated indentations produced forces of 21.21 ± 6.23 N (3D-printed depressor) and 25.02 ± 4.64 N (steel depressor), with no significant difference between devices. The 3D-printed instrument breakage point was 63.27 ± 10.72 N, with a 2.98 factor of safety (FS) and 3.39 reliability index (β). In the porcine model, scleral depression produced a 15.63 ± 8.13 mmHg increase in IOP, requiring 0.191 ± 0.09 N (FS = 331.2 and β = 5.88). The 3D-printed depressor demonstrates effective mechanical robustness, wide safety margins, and functional equivalence to steel instruments, supporting the use of customizable, low-cost 3D-printed depressors in training and clinical settings.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
University of Central Florida (US), Faculdade Evangélica do Paraná (BR), Instituto de Oncologia do Paraná (BR), Fundo Brasil (BR), Fundação Faculdade de Medicina (BR)
Fundação de Amparo à Pesquisa do Estado de São Paulo
Openalex Percentile: Top 10%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.