Analytical and Clinical Performance of the Atellica <scp>HEMA</scp> 580 Haematology Analyzer: A Comparison With the <scp>ADVIA</scp> 2120i

INTRODUCTION: The Atellica HEMA 580 is a novel haematology analyser providing eight-part leukocyte differential and optical platelet (PLT) counting. We compared its analytical and clinical performance with that of the ADVIA 2120i. METHODS: A total of 796 residual whole blood specimens were analysed. Carryover, linearity and precision were assessed. Inter-instrument comparison was performed for complete blood counts and five-part differential (5-Diff) parameters using Passing-Bablok regression and Bland-Altman analyses. Additional regression against manual microscopy was conducted for flagged samples, including PLT, reticulocytes and common 5-Diff parameters. RESULTS: ≥ 0.99). Precision was within acceptable limits; coefficient of variant was < 5% for most parameters, except optical PLT, lymphocytes, monocytes, eosinophils and nucleated red blood cells under specific conditions. Inter-instrument comparisons revealed remarkably high correlations (r ≥ 0.9) for most parameters, whereas mean corpuscular haemoglobin concentration (MCHC) (r = 0.429) and basophils (r = 0.627) showed weaker concordance. Passing-Bablok regression indicated minimal systematic or proportional bias for most parameters. Bland-Altman analysis showed that > 95% of results were within the 95% limits of agreement, except white blood cells, mean corpuscular volume and PLT (94.7%-94.8%). Comparison with microscopy confirmed good concordance for PLT and reticulocytes; however, lymphocytes and monocytes had proportional and systematic biases. CONCLUSION: The Atellica HEMA 580 demonstrated reliable analytical performance and strong agreement with the ADVIA 2120i. This supports its suitability as a routine haematology analyser, providing accurate and clinically comparable results.

Authors

Institutions

Publication Details

Journal
International Journal of Laboratory Hematology
Published
2026-04-10
DOI
https://doi.org/10.1111/ijlh.70107
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytical and Clinical Performance of the Atellica HEMA 580 Haematology Analyzer: A Comparison With the ADVIA 2120i

Gilsung Yoo, Juwon Kim, Hyeongyu Lee
International Journal of Laboratory Hematology
Clinical Laboratory Practices and Quality Control
article

Analytical and Clinical Performance of the Atellica HEMA 580 Haematology Analyzer: A Comparison With the ADVIA 2120i

Gilsung Yoo, Juwon Kim, Hyeongyu Lee
article en

Abstract

INTRODUCTION: The Atellica HEMA 580 is a novel haematology analyser providing eight-part leukocyte differential and optical platelet (PLT) counting. We compared its analytical and clinical performance with that of the ADVIA 2120i. METHODS: A total of 796 residual whole blood specimens were analysed. Carryover, linearity and precision were assessed. Inter-instrument comparison was performed for complete blood counts and five-part differential (5-Diff) parameters using Passing-Bablok regression and Bland-Altman analyses. Additional regression against manual microscopy was conducted for flagged samples, including PLT, reticulocytes and common 5-Diff parameters. RESULTS: ≥ 0.99). Precision was within acceptable limits; coefficient of variant was < 5% for most parameters, except optical PLT, lymphocytes, monocytes, eosinophils and nucleated red blood cells under specific conditions. Inter-instrument comparisons revealed remarkably high correlations (r ≥ 0.9) for most parameters, whereas mean corpuscular haemoglobin concentration (MCHC) (r = 0.429) and basophils (r = 0.627) showed weaker concordance. Passing-Bablok regression indicated minimal systematic or proportional bias for most parameters. Bland-Altman analysis showed that > 95% of results were within the 95% limits of agreement, except white blood cells, mean corpuscular volume and PLT (94.7%-94.8%). Comparison with microscopy confirmed good concordance for PLT and reticulocytes; however, lymphocytes and monocytes had proportional and systematic biases. CONCLUSION: The Atellica HEMA 580 demonstrated reliable analytical performance and strong agreement with the ADVIA 2120i. This supports its suitability as a routine haematology analyser, providing accurate and clinically comparable results.

International Journal of Laboratory Hematology
Yonsei University (KR)
Openalex Percentile: Top 17%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.