Biomechanical effects of a novel anatomic fusion cage for anterior cervical corpectomy: a finite element study

Anterior cervical corpectomy and fusion (ACCF) is a crucial surgical procedure for addressing cervical degenerative diseases. However, traditional titanium mesh cages (TTMC) are susceptible to complications, such as cage subsidence and internal fixation failure. In this study, we developed a novel anatomic titanium mesh cage (NATMC) and assessed its biomechanical properties through finite element analysis. A conventional three-dimensional finite element model comprising cervical vertebrae C2–C7 was established using computed tomography (CT) data. Subsequently, two surgical models (TTMC and NATMC) were generated based on the normal model. Uniform load conditions were applied to these distinct surgical models to simulate flexion, extension, lateral bending, and axial rotation of the cervical spine. The range of motion (ROM), stress peaks at the titanium mesh-endplate interface, screw–bone interface, and the adjacent intervertebral disc were recorded and compared between the two groups. In comparison to the intact model, the surgical segmental range of motion (ROM) for C4-6 was reduced by 96.89% in NATMC and 96.91% in TTMC. The stress peaks at the C4 inferior endplate and C6 superior endplate were higher in TTMC, measuring between 1.61 and 9.96 MPa and 1.46–14.29 MPa, respectively, followed by NATMC with values of 0.93–7.75 MPa and 1.08–11.69 MPa. Stress peaks on NATMC were lower (68.47–309.90 MPa) compared to TTMC (86.80–368.30 MPa). Regarding stress peaks at the screw-bone interface, NATMC exhibited lower values (11.91–42.49 MPa) compared to TTMC (35.49–153.00 MPa). Finally, comparing stress peaks in adjacent discs (C3/4 and C6/7), NATMC and TTMC showed similar results, ranging between 1.56 and 5.47 MPa in C3/4 and 1.56–2.91 MPa in C6/7, respectively. The novel anatomic titanium mesh cage demonstrated effective reduction in vertebral subsidence and minimized pressure on the internal fixation interface following anterior cervical corpectomy and fusion (ACCF). It demonstrated superior stability and safety compared to traditional titanium mesh cage.

Authors

Institutions

Publication Details

Journal
BMC Musculoskeletal Disorders
Published
2026-04-11
DOI
https://doi.org/10.1186/s12891-026-09814-w
Primary Topic
Cervical and Thoracic Myelopathy
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanical effects of a novel anatomic fusion cage for anterior cervical corpectomy: a finite element study

Ze Lin, Bo Wei, Zhijun Li, Biru Liang et al.
BMC Musculoskeletal Disorders
Cervical and Thoracic Myelopathy
article

Biomechanical effects of a novel anatomic fusion cage for anterior cervical corpectomy: a finite element study

Ze Lin, Bo Wei, Zhijun Li, Biru Liang, Yue He, Wenhua Huang, Yuzhu Wang, Zhanglin Wu, Hanbin Ouyang
article en

Abstract

Anterior cervical corpectomy and fusion (ACCF) is a crucial surgical procedure for addressing cervical degenerative diseases. However, traditional titanium mesh cages (TTMC) are susceptible to complications, such as cage subsidence and internal fixation failure. In this study, we developed a novel anatomic titanium mesh cage (NATMC) and assessed its biomechanical properties through finite element analysis. A conventional three-dimensional finite element model comprising cervical vertebrae C2–C7 was established using computed tomography (CT) data. Subsequently, two surgical models (TTMC and NATMC) were generated based on the normal model. Uniform load conditions were applied to these distinct surgical models to simulate flexion, extension, lateral bending, and axial rotation of the cervical spine. The range of motion (ROM), stress peaks at the titanium mesh-endplate interface, screw–bone interface, and the adjacent intervertebral disc were recorded and compared between the two groups. In comparison to the intact model, the surgical segmental range of motion (ROM) for C4-6 was reduced by 96.89% in NATMC and 96.91% in TTMC. The stress peaks at the C4 inferior endplate and C6 superior endplate were higher in TTMC, measuring between 1.61 and 9.96 MPa and 1.46–14.29 MPa, respectively, followed by NATMC with values of 0.93–7.75 MPa and 1.08–11.69 MPa. Stress peaks on NATMC were lower (68.47–309.90 MPa) compared to TTMC (86.80–368.30 MPa). Regarding stress peaks at the screw-bone interface, NATMC exhibited lower values (11.91–42.49 MPa) compared to TTMC (35.49–153.00 MPa). Finally, comparing stress peaks in adjacent discs (C3/4 and C6/7), NATMC and TTMC showed similar results, ranging between 1.56 and 5.47 MPa in C3/4 and 1.56–2.91 MPa in C6/7, respectively. The novel anatomic titanium mesh cage demonstrated effective reduction in vertebral subsidence and minimized pressure on the internal fixation interface following anterior cervical corpectomy and fusion (ACCF). It demonstrated superior stability and safety compared to traditional titanium mesh cage.

BMC Musculoskeletal Disorders
Guangdong Medical College (CN), Guangzhou University (CN), Third Affiliated Hospital of Southern Medical University (CN), Zhongshan Hospital (CN), Community Health Center (US), Affiliated Hospital of Guangdong Medical College Hospital (CN), Southern Medical University (CN)
National Natural Science Foundation of China, National Key Research and Development Program of China, Basic and Applied Basic Research Foundation of Guangdong Province
Openalex Percentile: Top 6%
Cervical and Thoracic Myelopathy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.