Lipid Interference Flagging and Haemoglobin Correction in Lipemic Samples With a Hematology Analyzer

BACKGROUND: Lipemia can interfere with haemoglobin (HGB) levels, which is critical for evaluating anemia and polycythemia. This study aims to develop and evaluate the performance of a lipid flagging and haemoglobin correction system. METHODS: Lipemic samples were simulated by the addition of a fat emulsion to 79 whole-blood samples from healthy participants. Additional blood samples were collected from 152 patients with lipemia. Plasma exchange was performed to eliminate lipid interference and establish reference HGB values. A lipid interference flagging method was developed, showing the "Lipid Particles?" flag on the hematology analyzer. Lipid interference was defined as samples showing a percentage change in HGB (ΔHGB%) before and after plasma exchange exceeding 4%. The accuracy of the optical haemoglobin (HGB-O) value was assessed in the samples with confirmed lipid interference. RESULTS: A flagging method was developed based on the changes in routine blood parameters and blood cell scattergram. Among the 152 clinical lipemic samples, 27 were identified as lipid interference. While the "Lipid Particles?" flag was triggered for 28 samples, yielding a sensitivity of 92.59%, specificity of 97.60%, positive predictive value (PPV) of 89.29%, and negative predictive value (NPV) of 98.39%. Additionally, among the 27 lipid-interfered samples, the HGB-O value showed excellent correlation (Pearson's r = 0.935) and an average deviation of 0.72% (95% CI, -1.02% to 2.47%) with the reference value. CONCLUSIONS: These findings highlight the clinical value of the "Lipid Particles?" flag in identifying lipid interference, demonstrating the accuracy of HGB-O in providing a reliable haemoglobin measurement.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Laboratory Analysis
Published
2026-04-10
DOI
https://doi.org/10.1002/jcla.70223
Primary Topic
Diabetes, Cardiovascular Risks, and Lipoproteins
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Lipid Interference Flagging and Haemoglobin Correction in Lipemic Samples With a Hematology Analyzer

Binhua Zhao, Yushuai Ma, Z. Chen, Yinhua Fan et al.
Journal of Clinical Laboratory Analysis
Diabetes, Cardiovascular Risks, and Lipoproteins
article

Lipid Interference Flagging and Haemoglobin Correction in Lipemic Samples With a Hematology Analyzer

Binhua Zhao, Yushuai Ma, Z. Chen, Yinhua Fan, Yucheng Shen, Jun Cao, Yong Wang
article en

Abstract

BACKGROUND: Lipemia can interfere with haemoglobin (HGB) levels, which is critical for evaluating anemia and polycythemia. This study aims to develop and evaluate the performance of a lipid flagging and haemoglobin correction system. METHODS: Lipemic samples were simulated by the addition of a fat emulsion to 79 whole-blood samples from healthy participants. Additional blood samples were collected from 152 patients with lipemia. Plasma exchange was performed to eliminate lipid interference and establish reference HGB values. A lipid interference flagging method was developed, showing the "Lipid Particles?" flag on the hematology analyzer. Lipid interference was defined as samples showing a percentage change in HGB (ΔHGB%) before and after plasma exchange exceeding 4%. The accuracy of the optical haemoglobin (HGB-O) value was assessed in the samples with confirmed lipid interference. RESULTS: A flagging method was developed based on the changes in routine blood parameters and blood cell scattergram. Among the 152 clinical lipemic samples, 27 were identified as lipid interference. While the "Lipid Particles?" flag was triggered for 28 samples, yielding a sensitivity of 92.59%, specificity of 97.60%, positive predictive value (PPV) of 89.29%, and negative predictive value (NPV) of 98.39%. Additionally, among the 27 lipid-interfered samples, the HGB-O value showed excellent correlation (Pearson's r = 0.935) and an average deviation of 0.72% (95% CI, -1.02% to 2.47%) with the reference value. CONCLUSIONS: These findings highlight the clinical value of the "Lipid Particles?" flag in identifying lipid interference, demonstrating the accuracy of HGB-O in providing a reliable haemoglobin measurement.

Journal of Clinical Laboratory Analysis
Soochow University (CN), First Affiliated Hospital of Soochow University (CN), Mindray (China) (CN)
Openalex Percentile: Top 20%
Diabetes, Cardiovascular Risks, and Lipoproteins
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Lipid Interference Flagging and Haemoglobin Correction in Lipemic Samples With a Hematology Analyzer — Binhua Zhao, Yushuai Ma, et al. · Journal of Clinical Laboratory Analysis (2026) | TGRS Research Map | TGRS