Analytical and clinical performance of the Flash10 molecular point-of-care system for detection of influenza A and B viruses

OBJECTIVES: To evaluate the analytical performance and clinical workflow efficiency of the Flash10 molecular point-of-care testing (POCT) system for the detection of influenza A (FluA) and influenza B (FluB) viruses. METHODS: Throat swab specimens from 178 individuals presenting with influenza-like symptoms were tested in parallel using the Flash10 system, GeneXpert, and real-time quantitative PCR (RT-PCR). Analytical performance was assessed in terms of precision, limit of detection (LOD), specificity, and contamination risk. Result agreement among platforms was evaluated using correlation analysis, Bland-Altman analysis, and Kappa statistics. Turnaround time (TAT) was retrieved from the laboratory information system to compare clinical workflow efficiency between Flash10 and conventional RT-PCR. RESULTS: The Flash10 system demonstrated good precision, with intra-batch coefficients of variation (CVs) ranging from 0.79 % to 1.41 % and inter-batch CVs ranging from 2.62 % to 3.05 % at concentrations of 500 and 1,000 copies/mL. The LOD for both FluA and FluB was 500 copies/mL, with a positive detection rate of 100 %. No cross-reactivity was observed with 10 common respiratory pathogens, and no contamination was detected during testing. Ct values obtained by Flash10 showed strong correlations with those generated by RT-PCR and GeneXpert for both FluA (r=0.87-0.93, p<0.0001) and FluB (rs=0.86-0.90, p<0.0001). Bland-Altman analysis revealed consistent agreement, despite systematic Ct differences between methods. Qualitative agreement for positive and negative results was perfect (Kappa=1.000, p<0.0001). The median TAT for Flash10 was 66.0 min, significantly shorter than that of RT-PCR (1,184.0 min, p<0.0001). CONCLUSIONS: The Flash10 molecular POCT system demonstrates analytical performance and result consistency comparable to established molecular diagnostic methods for the detection of FluA and FluB viruses. Its substantially shorter turnaround time supports its potential utility for rapid molecular testing in clinical point-of-care settings.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-04-11
DOI
https://doi.org/10.1515/cclm-2026-0093
Primary Topic
Influenza Virus Research Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytical and clinical performance of the Flash10 molecular point-of-care system for detection of influenza A and B viruses

Sheng Zeng, Xiaoling Chen, Pei Yan, Huidi Peng et al.
Clinical Chemistry and Laboratory Medicine (CCLM)
Influenza Virus Research Studies
article

Analytical and clinical performance of the Flash10 molecular point-of-care system for detection of influenza A and B viruses

Sheng Zeng, Xiaoling Chen, Pei Yan, Huidi Peng, Shengjin Cui, Yanting Li, Yan Li, Xiaoping Xu, Yuling Zeng
article en

Abstract

OBJECTIVES: To evaluate the analytical performance and clinical workflow efficiency of the Flash10 molecular point-of-care testing (POCT) system for the detection of influenza A (FluA) and influenza B (FluB) viruses. METHODS: Throat swab specimens from 178 individuals presenting with influenza-like symptoms were tested in parallel using the Flash10 system, GeneXpert, and real-time quantitative PCR (RT-PCR). Analytical performance was assessed in terms of precision, limit of detection (LOD), specificity, and contamination risk. Result agreement among platforms was evaluated using correlation analysis, Bland-Altman analysis, and Kappa statistics. Turnaround time (TAT) was retrieved from the laboratory information system to compare clinical workflow efficiency between Flash10 and conventional RT-PCR. RESULTS: The Flash10 system demonstrated good precision, with intra-batch coefficients of variation (CVs) ranging from 0.79 % to 1.41 % and inter-batch CVs ranging from 2.62 % to 3.05 % at concentrations of 500 and 1,000 copies/mL. The LOD for both FluA and FluB was 500 copies/mL, with a positive detection rate of 100 %. No cross-reactivity was observed with 10 common respiratory pathogens, and no contamination was detected during testing. Ct values obtained by Flash10 showed strong correlations with those generated by RT-PCR and GeneXpert for both FluA (r=0.87-0.93, p<0.0001) and FluB (rs=0.86-0.90, p<0.0001). Bland-Altman analysis revealed consistent agreement, despite systematic Ct differences between methods. Qualitative agreement for positive and negative results was perfect (Kappa=1.000, p<0.0001). The median TAT for Flash10 was 66.0 min, significantly shorter than that of RT-PCR (1,184.0 min, p<0.0001). CONCLUSIONS: The Flash10 molecular POCT system demonstrates analytical performance and result consistency comparable to established molecular diagnostic methods for the detection of FluA and FluB viruses. Its substantially shorter turnaround time supports its potential utility for rapid molecular testing in clinical point-of-care settings.

Clinical Chemistry and Laboratory Medicine (CCLM)
University of Hong Kong - Shenzhen Hospital (CN), University of Hong Kong (HK)
Openalex Percentile: Top 17%
Influenza Virus Research Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.