Assessment of LD‐V1 radiochromic film low‐dose performance for mega‐voltage radiotherapy quality assurance

BACKGROUND: Highly modulated radiotherapy delivery demands functional multileaf collimator (MLC) quality assurance (QA) with the ability to measure very low doses during short (<1 s) temporal segments of beam delivery. A radiochromic film with adequate sensitivity in the few-cGy range is therefore needed. LD-V1 film was designed for 2-20 cGy measurements in diagnostic kV energy ranges, but its performance for megavoltage (MV) beam qualities has not been well characterized. PURPOSE: To demonstrate the feasibility of LD-V1 film for low-dose MV measurements and characterize its signal dependence under conditions relevant for clinical QA. METHODS: A 6 MV calibration curve was generated using eight dose levels between 0 and 40 cGy and compared with EBT4 film irradiated under identical conditions. Film characteristics evaluated included dose response, scanner and film signal uniformity, scan repeatability, energy dependence, post-exposure signal evolution, scanning-orientation dependence, and one-dimensional (1-D) profile visualization. RESULTS: LD-V1 demonstrated higher net optical density (OD) and significantly less measurement uncertainty compared to EBT4 at dose levels less than 40 cGy. Significant energy dependence between kV and MV beam qualities was seen. Average net OD differences of 6.3% were seen between landscape and portrait scanning orientations. Post-irradiation signal growth followed a logarithmic trend, stabilizing at ∼24 h. Scan-to-scan variation was <0.1%. Film-signal uniformity was within 3.9%. 1-D profiles showed discernible dose fall-off for doses as low as 10 cGy. CONCLUSIONS: LD-V1 demonstrates suitability for low-dose MV measurements and exhibits improved low-dose sensitivity relative to EBT4, indicating its potential value for MLC QA in highly dynamic treatment deliveries. Nonetheless, its pronounced energy and scanning-orientation dependence underscores the need for rigorous and consistent calibration and handling procedures.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-04-01
DOI
https://doi.org/10.1002/acm2.70578
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of LD‐V1 radiochromic film low‐dose performance for mega‐voltage radiotherapy quality assurance

Victor Malkov, J. John Lucido, James Kavanaugh, D. Garcia et al.
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Assessment of LD‐V1 radiochromic film low‐dose performance for mega‐voltage radiotherapy quality assurance

Victor Malkov, J. John Lucido, James Kavanaugh, D. Garcia, Ruiming C. Edmondson
article en

Abstract

BACKGROUND: Highly modulated radiotherapy delivery demands functional multileaf collimator (MLC) quality assurance (QA) with the ability to measure very low doses during short (<1 s) temporal segments of beam delivery. A radiochromic film with adequate sensitivity in the few-cGy range is therefore needed. LD-V1 film was designed for 2-20 cGy measurements in diagnostic kV energy ranges, but its performance for megavoltage (MV) beam qualities has not been well characterized. PURPOSE: To demonstrate the feasibility of LD-V1 film for low-dose MV measurements and characterize its signal dependence under conditions relevant for clinical QA. METHODS: A 6 MV calibration curve was generated using eight dose levels between 0 and 40 cGy and compared with EBT4 film irradiated under identical conditions. Film characteristics evaluated included dose response, scanner and film signal uniformity, scan repeatability, energy dependence, post-exposure signal evolution, scanning-orientation dependence, and one-dimensional (1-D) profile visualization. RESULTS: LD-V1 demonstrated higher net optical density (OD) and significantly less measurement uncertainty compared to EBT4 at dose levels less than 40 cGy. Significant energy dependence between kV and MV beam qualities was seen. Average net OD differences of 6.3% were seen between landscape and portrait scanning orientations. Post-irradiation signal growth followed a logarithmic trend, stabilizing at ∼24 h. Scan-to-scan variation was <0.1%. Film-signal uniformity was within 3.9%. 1-D profiles showed discernible dose fall-off for doses as low as 10 cGy. CONCLUSIONS: LD-V1 demonstrates suitability for low-dose MV measurements and exhibits improved low-dose sensitivity relative to EBT4, indicating its potential value for MLC QA in highly dynamic treatment deliveries. Nonetheless, its pronounced energy and scanning-orientation dependence underscores the need for rigorous and consistent calibration and handling procedures.

Journal of Applied Clinical Medical PhysicsVol. 27(4)
Mayo Clinic (US)
Affordable and clean energy
Openalex Percentile: Top 18%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.