On the stability of CMOS ICs with plasma-enhanced silicone encapsulation for active implantable neurotechnology: 4.3 years of accelerated life testing

The long-term stability of polymer-encapsulated CMOS integrated circuits (ICs) is essential for mm-size active implantable medical devices (AIMDs), where hermetic packaging is impractical. Validating robust, biocompatible, implantable IC encapsulation is a prerequisite for chip-scale polymeric bioelectronic implants. This study presents the first long-term validation of a design strategy combining medical-grade silicone, plasma surface treatment for adhesion, and foundry-manufactured CMOS ICs. These ICs feature a perforated shield layer (top metal layer) and a double-layer wall-of-vias to reduce moisture ingress and mechanical delamination. Test structures with silicon oxide/nitride passivation were encapsulated using implant-compatible processes and immersed in saline under accelerated ageing conditions (47 °C, 67 °C, and 87 °C) for up to 4.3 years, under DC and biphasic biases. Throughout the study, electrochemical impedance spectroscopy (EIS) showed no insulation failures. Minor visual corrosion was confined to wire bonds and solder pads, with no correlation to electrical degradation. These results demonstrate the robustness to biofluid exposure of modern IC passivation when combined with well-adhered silicone encapsulation. To our knowledge, this is the longest and most comprehensive accelerated ageing study of its kind, and the first to establish a scalable, industry-compatible encapsulation method for implanted ICs. Our findings provide critical evidence supporting the integration of CMOS ICs into next-generation bioelectronic implants.

Authors

Institutions

Publication Details

Journal
Frontiers in Electronics
Published
2026-04-07
DOI
https://doi.org/10.3389/felec.2026.1788559
Primary Topic
Neuroscience and Neural Engineering
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

On the stability of CMOS ICs with plasma-enhanced silicone encapsulation for active implantable neurotechnology: 4.3 years of accelerated life testing

Callum Lamont, Kambiz Nanbakhsh, Anne Vanhoestenberghe, Ahmad Shah Idil et al.
Frontiers in Electronics
Neuroscience and Neural Engineering
article

On the stability of CMOS ICs with plasma-enhanced silicone encapsulation for active implantable neurotechnology: 4.3 years of accelerated life testing

Callum Lamont, Kambiz Nanbakhsh, Anne Vanhoestenberghe, Ahmad Shah Idil, Federico Mazza, Vasiliki Giagka, Nicholas de Neufville Donaldson, Timothy G. Constandinou
article en

Abstract

The long-term stability of polymer-encapsulated CMOS integrated circuits (ICs) is essential for mm-size active implantable medical devices (AIMDs), where hermetic packaging is impractical. Validating robust, biocompatible, implantable IC encapsulation is a prerequisite for chip-scale polymeric bioelectronic implants. This study presents the first long-term validation of a design strategy combining medical-grade silicone, plasma surface treatment for adhesion, and foundry-manufactured CMOS ICs. These ICs feature a perforated shield layer (top metal layer) and a double-layer wall-of-vias to reduce moisture ingress and mechanical delamination. Test structures with silicon oxide/nitride passivation were encapsulated using implant-compatible processes and immersed in saline under accelerated ageing conditions (47 °C, 67 °C, and 87 °C) for up to 4.3 years, under DC and biphasic biases. Throughout the study, electrochemical impedance spectroscopy (EIS) showed no insulation failures. Minor visual corrosion was confined to wire bonds and solder pads, with no correlation to electrical degradation. These results demonstrate the robustness to biofluid exposure of modern IC passivation when combined with well-adhered silicone encapsulation. To our knowledge, this is the longest and most comprehensive accelerated ageing study of its kind, and the first to establish a scalable, industry-compatible encapsulation method for implanted ICs. Our findings provide critical evidence supporting the integration of CMOS ICs into next-generation bioelectronic implants.

Frontiers in ElectronicsVol. 7
King's College London (GB), Great Lakes NeuroTechnologies (United States) (US), UK Dementia Research Institute (GB), Fraunhofer Institute for Reliability and Microintegration (DE), University College London (GB), Imperial College London (GB), Delft University of Technology (NL)
Wellcome Trust, European Commission, Electronic Components and Systems for European Leadership, General Sir John Monash Foundation, Horizon 2020 Framework Programme, Engineering and Physical Sciences Research Council
Openalex Percentile: Top 29%
Neuroscience and Neural Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.