Design and Characterization of a Low-Current Compact Extraction and Matching System for a High-Frequency RFQ in Medical Applications

A low-current compact extraction and matching system has been designed and experimentally tested to evaluate its capability for direct proton injection of 15 keV into a 750 MHz radiofrequency quadrupole for medical applications. The design methodology combined 2D and 3D layouts, supported by detailed electrostatic simulations. First experimental results are reported, including beam current characterization and irradiation measurements under varying operating conditions. These results are benchmarked against simulation data to provide a preliminary evaluation of system performance. Ongoing efforts focus on extending the experimental campaign to consolidate these findings. A comparative study with a gridded-lens-based extraction and transport system is also being conducted to achieve reliable matching of high-quality beams to high-frequency RFQs for clinical implementation.

Authors

Institutions

Publication Details

Journal
Instruments
Published
2026-04-06
DOI
https://doi.org/10.3390/instruments10020022
Primary Topic
Particle accelerators and beam dynamics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and Characterization of a Low-Current Compact Extraction and Matching System for a High-Frequency RFQ in Medical Applications

A. Mamaras, A. Lombardi, Francesco Di Lorenzo, Eleonora Pasino et al.
Instruments
Particle accelerators and beam dynamics
article

Design and Characterization of a Low-Current Compact Extraction and Matching System for a High-Frequency RFQ in Medical Applications

A. Mamaras, A. Lombardi, Francesco Di Lorenzo, Eleonora Pasino, Dimitrios Sampsonidis
article en

Abstract

A low-current compact extraction and matching system has been designed and experimentally tested to evaluate its capability for direct proton injection of 15 keV into a 750 MHz radiofrequency quadrupole for medical applications. The design methodology combined 2D and 3D layouts, supported by detailed electrostatic simulations. First experimental results are reported, including beam current characterization and irradiation measurements under varying operating conditions. These results are benchmarked against simulation data to provide a preliminary evaluation of system performance. Ongoing efforts focus on extending the experimental campaign to consolidate these findings. A comparative study with a gridded-lens-based extraction and transport system is also being conducted to achieve reliable matching of high-quality beams to high-frequency RFQs for clinical implementation.

InstrumentsVol. 10(2)
Aristotle University of Thessaloniki (GR), European Organization for Nuclear Research (CH)
Openalex Percentile: Top 8%
Particle accelerators and beam dynamics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.