Static and Dynamic Compliance of Thick-Walled Vessel Models: Mathematical Modeling and Experimental Validation Using Silicone and Polyvinyl Alcohol (PVA) Hydrogel

PURPOSE: Thick-walled vascular models, including block models, are increasingly used for in vitro investigations of fluid dynamics and endovascular device testing, particularly fatigue testing at high frequencies, as valuable alternatives to thin-walled models due to their simpler fabrication. Physiologically compliant models provide more realistic insights into hemodynamics and implant deformation; however, no compact and experimentally validated closed-form equation exists to predict the volumetric compliance of thick-walled vessel models. METHODS: A closed-form mathematical equation was derived to estimate the volumetric compliance of thick-walled cylindrical vessel models. The equation was experimentally validated through static compliance tests on cylindrical vascular models of varying wall thicknesses fabricated from two silicone elastomers and polyvinyl alcohol hydrogel (PVA-H) with four PVA concentrations (10, 12, 14, and 16 wt/vol.%). Dynamic compliance tests were performed at 1, 5, 10, and 30 Hz to assess the viscoelastic behavior of the materials. Finally, an applicability study was performed by fabricating block-form anatomical aneurysm models targeting physiological volumetric compliance based on the predicted equation. RESULTS: > 0.95). PVA-H-10 exhibited volumetric compliance within the physiological range (0.4-1.1%/mmHg). Volumetric compliance decreased by over 50% at 30 Hz, confirming viscoelastic behavior with a maximum loss to storage volumetric compliance ratio of 0.14. Measured aneurysm model compliance matched predictions with an error below 7%. CONCLUSION: This study provided a validated mathematical and experimental framework for fabricating thick-walled vascular models with physiologically relevant volumetric compliance and viscoelasticity for fluid dynamics and endovascular device testing.

Authors

Institutions

Publication Details

Journal
Annals of Biomedical Engineering
Published
2026-04-07
DOI
https://doi.org/10.1007/s10439-026-04130-9
Primary Topic
Aortic aneurysm repair treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Static and Dynamic Compliance of Thick-Walled Vessel Models: Mathematical Modeling and Experimental Validation Using Silicone and Polyvinyl Alcohol (PVA) Hydrogel

Giorgio Cattaneo, Ashkan Shiravand, Wasilios Hatziklitiu, Fabian Baumer et al.
Annals of Biomedical Engineering
Aortic aneurysm repair treatments
article

Static and Dynamic Compliance of Thick-Walled Vessel Models: Mathematical Modeling and Experimental Validation Using Silicone and Polyvinyl Alcohol (PVA) Hydrogel

Giorgio Cattaneo, Ashkan Shiravand, Wasilios Hatziklitiu, Fabian Baumer, Elisabeth Janzen
article en

Abstract

PURPOSE: Thick-walled vascular models, including block models, are increasingly used for in vitro investigations of fluid dynamics and endovascular device testing, particularly fatigue testing at high frequencies, as valuable alternatives to thin-walled models due to their simpler fabrication. Physiologically compliant models provide more realistic insights into hemodynamics and implant deformation; however, no compact and experimentally validated closed-form equation exists to predict the volumetric compliance of thick-walled vessel models. METHODS: A closed-form mathematical equation was derived to estimate the volumetric compliance of thick-walled cylindrical vessel models. The equation was experimentally validated through static compliance tests on cylindrical vascular models of varying wall thicknesses fabricated from two silicone elastomers and polyvinyl alcohol hydrogel (PVA-H) with four PVA concentrations (10, 12, 14, and 16 wt/vol.%). Dynamic compliance tests were performed at 1, 5, 10, and 30 Hz to assess the viscoelastic behavior of the materials. Finally, an applicability study was performed by fabricating block-form anatomical aneurysm models targeting physiological volumetric compliance based on the predicted equation. RESULTS: > 0.95). PVA-H-10 exhibited volumetric compliance within the physiological range (0.4-1.1%/mmHg). Volumetric compliance decreased by over 50% at 30 Hz, confirming viscoelastic behavior with a maximum loss to storage volumetric compliance ratio of 0.14. Measured aneurysm model compliance matched predictions with an error below 7%. CONCLUSION: This study provided a validated mathematical and experimental framework for fabricating thick-walled vascular models with physiologically relevant volumetric compliance and viscoelasticity for fluid dynamics and endovascular device testing.

Annals of Biomedical Engineering
University of Stuttgart (DE), Stuttgart Technical University of Applied Sciences (DE)
Universität Stuttgart
Openalex Percentile: Top 23%
Aortic aneurysm repair treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.