Automated contrast-to-noise ratio analysis in chest CT: validation of an open-source segmentation approach

OBJECTIVES: This study aimed to evaluate the feasibility and accuracy of automated contrast-to-noise ratio (CNR) analysis in chest CT using the open-source body and organ analysis (BOA) framework and to validate segmentation modifications for reproducible image-quality assessment. MATERIALS AND METHODS: This retrospective study analyzed 100 contrast-enhanced chest CTs (mean age 60.2 ± 15 years; 40% female; 50 CTA, 50 CTPA) and validated the approach in an external cancer imaging archive (TCIA) cohort (n = 100). Automated BOA segmentations of the aorta, pulmonary trunk, and paraspinal muscles were modified by fat subtraction and binary erosion and compared with manual measurements from three radiologists. Agreement was assessed using statistical testing, Bland-Altman analysis, and intraclass correlation coefficients (ICC). RESULTS: Unmodified BOA segmentations yielded significantly lower CNRs than manual measurements (all p < 0.01, mean difference up to 6.3). Fat subtraction and binary erosion progressively reduced deviations, with the optimized variant (m_erode6 combined with a_erode6 or p_erode6) showing no significant differences from radiologists (p ≥ 0.35). In the external TCIA validation cohort (n = 100), agreement was excellent (ICC 0.89-0.93), and Bland-Altman analysis demonstrated minimal bias (Aorta: 0.16 [limits of agreement (LoA) -3.47 to 3.80]; PT: 0.42 [LoA -4.03 to 4.87]). CONCLUSIONS: A minimally modified open-source segmentation framework enables fully automated, reproducible CNR assessment in chest CT, achieving expert-level agreement, including robust performance in external validation. This scalable alternative to manual region-of-interest (ROI) measurement streamlines image-quality assessment, facilitates protocol optimization, and provides standardized metrics ready for integration into AI workflows. CRITICAL RELEVANCE STATEMENT: This study provides a validated, fully automated method for quantitative CT image quality assessment, reducing observer dependence and enabling consistent evaluation across scanners, protocols, and institutions, thereby supporting reproducible image quality metrics in clinical routine. KEY POINTS: Automated CNR assessment enables objective and reproducible evaluation of image quality in CTA and CTPA. Adjustments of the segmentation strategy can substantially improve the accuracy of automated measurements. The fully automated approach provides a foundation for standardized and scalable CT image quality analysis in research and clinical practice.

Authors

Institutions

Publication Details

Journal
Insights into Imaging
Published
2026-04-07
DOI
https://doi.org/10.1186/s13244-026-02263-y
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated contrast-to-noise ratio analysis in chest CT: validation of an open-source segmentation approach

Luca Salhöfer, René Hosch, Nikolas Beck, Lale Umutlu et al.
Insights into Imaging
Cardiac Imaging and Diagnostics
article

Automated contrast-to-noise ratio analysis in chest CT: validation of an open-source segmentation approach

Luca Salhöfer, René Hosch, Nikolas Beck, Lale Umutlu, Denise Bos, Giulia Baldini, Sebastian Zensen, Felix Nensa, Mathias Holtkamp, Jannis Straus, J. Haubold, Michael Forsting, Marcel Opitz
article en

Abstract

OBJECTIVES: This study aimed to evaluate the feasibility and accuracy of automated contrast-to-noise ratio (CNR) analysis in chest CT using the open-source body and organ analysis (BOA) framework and to validate segmentation modifications for reproducible image-quality assessment. MATERIALS AND METHODS: This retrospective study analyzed 100 contrast-enhanced chest CTs (mean age 60.2 ± 15 years; 40% female; 50 CTA, 50 CTPA) and validated the approach in an external cancer imaging archive (TCIA) cohort (n = 100). Automated BOA segmentations of the aorta, pulmonary trunk, and paraspinal muscles were modified by fat subtraction and binary erosion and compared with manual measurements from three radiologists. Agreement was assessed using statistical testing, Bland-Altman analysis, and intraclass correlation coefficients (ICC). RESULTS: Unmodified BOA segmentations yielded significantly lower CNRs than manual measurements (all p < 0.01, mean difference up to 6.3). Fat subtraction and binary erosion progressively reduced deviations, with the optimized variant (m_erode6 combined with a_erode6 or p_erode6) showing no significant differences from radiologists (p ≥ 0.35). In the external TCIA validation cohort (n = 100), agreement was excellent (ICC 0.89-0.93), and Bland-Altman analysis demonstrated minimal bias (Aorta: 0.16 [limits of agreement (LoA) -3.47 to 3.80]; PT: 0.42 [LoA -4.03 to 4.87]). CONCLUSIONS: A minimally modified open-source segmentation framework enables fully automated, reproducible CNR assessment in chest CT, achieving expert-level agreement, including robust performance in external validation. This scalable alternative to manual region-of-interest (ROI) measurement streamlines image-quality assessment, facilitates protocol optimization, and provides standardized metrics ready for integration into AI workflows. CRITICAL RELEVANCE STATEMENT: This study provides a validated, fully automated method for quantitative CT image quality assessment, reducing observer dependence and enabling consistent evaluation across scanners, protocols, and institutions, thereby supporting reproducible image quality metrics in clinical routine. KEY POINTS: Automated CNR assessment enables objective and reproducible evaluation of image quality in CTA and CTPA. Adjustments of the segmentation strategy can substantially improve the accuracy of automated measurements. The fully automated approach provides a foundation for standardized and scalable CT image quality analysis in research and clinical practice.

Insights into ImagingVol. 17(1)
University Hospital of Zurich (CH), Essen University Hospital (DE)
Deutsche Forschungsgemeinschaft
Openalex Percentile: Top 26%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies