Apparent pH Sensitivity of Non-Functionalized Graphene FETs Originates from Faradaic Reactions at the Gold Gate

Abstract Graphene field-effect transistors (GFETs) with solid-state gold gate electrodes have recently been used for accurate chemical analysis in complex matrices, yet the mechanistic origin of their pH sensitivity is not fully understood. Here we show that replacing only the gate electrode transforms an otherwise pH-insensitive graphene device into a pH-sensitive one. Using the same high-quality, low-defect CVD graphene as the channel, devices gated with a conventional Ag|AgCl reference electrode exhibit negligible pH sensitivity, while gold-gated devices show a response of −62.7 mV pH−1 at a gate voltage scan rate of 0.3 V s−1. Chronoamperometry distinguishes the transient charging current from the subsequent Faradaic contribution. Cyclic voltammetry and scan-rate-dependent measurements support a coupled Faradaic mechanism. The gate leakage current scales with the square root of the gate voltage scan rate (v), whereas the Dirac-point voltage varies approximately linearly with v1/2 and has a non-zero intercept. Independent open-circuit potential (OCP) measurements confirm that the Au gate itself is pH sensitive and relaxes slowly toward a stable potential. The GFET sensitivity decreases in magnitude at lower scan rates, and extrapolation as the scan rate approaches zero brings it closer to the OCP sensitivity. The remaining difference may reflect the different electrochemical conditions at the Au gate under bias and at open circuit. These findings indicate that the apparent pH response in Au-gated GFETs originates from gate-electrode electrochemistry and is read out through field-effect modulation of the graphene channel.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2026-11-14
DOI
https://doi.org/10.1021/acs.analchem.6c03272
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Apparent pH Sensitivity of Non-Functionalized Graphene FETs Originates from Faradaic Reactions at the Gold Gate

Kunshu Wang, Pengfei Lian, Zebin Kong, Shuo Wang et al.
Analytical Chemistry
Analytical Chemistry and Sensors
article

Apparent pH Sensitivity of Non-Functionalized Graphene FETs Originates from Faradaic Reactions at the Gold Gate

Kunshu Wang, Pengfei Lian, Zebin Kong, Shuo Wang, Bo WANG, Han Chen, Min Zhang
article en

Abstract

Abstract Graphene field-effect transistors (GFETs) with solid-state gold gate electrodes have recently been used for accurate chemical analysis in complex matrices, yet the mechanistic origin of their pH sensitivity is not fully understood. Here we show that replacing only the gate electrode transforms an otherwise pH-insensitive graphene device into a pH-sensitive one. Using the same high-quality, low-defect CVD graphene as the channel, devices gated with a conventional Ag|AgCl reference electrode exhibit negligible pH sensitivity, while gold-gated devices show a response of −62.7 mV pH−1 at a gate voltage scan rate of 0.3 V s−1. Chronoamperometry distinguishes the transient charging current from the subsequent Faradaic contribution. Cyclic voltammetry and scan-rate-dependent measurements support a coupled Faradaic mechanism. The gate leakage current scales with the square root of the gate voltage scan rate (v), whereas the Dirac-point voltage varies approximately linearly with v1/2 and has a non-zero intercept. Independent open-circuit potential (OCP) measurements confirm that the Au gate itself is pH sensitive and relaxes slowly toward a stable potential. The GFET sensitivity decreases in magnitude at lower scan rates, and extrapolation as the scan rate approaches zero brings it closer to the OCP sensitivity. The remaining difference may reflect the different electrochemical conditions at the Au gate under bias and at open circuit. These findings indicate that the apparent pH response in Au-gated GFETs originates from gate-electrode electrochemistry and is read out through field-effect modulation of the graphene channel.

Analytical Chemistry
Shanghai Jiao Tong University (CN), East China Normal University (CN), Shandong University of Science and Technology (CN)
Analytical Chemistry and Sensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon