Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein Gaps

Background: Pulsed field ablation (PFA) for atrial fibrillation (AF) requires adequate catheter-tissue contact. The Contact Index visualizes impedance change for the pentaspline catheter, but a validated contact threshold is unknown. Objective: The objective of this study is to define an impedance-change cutoff that predicts sufficient contact to prevent acute pulmonary vein gaps (PVG) during pulmonary vein isolation (PVI), and to validate Contact-Index-guided PVI. Methods: = 26) delivered energy only when all target splines met/exceeded the cutoff value obtained from Phase 1. Results: < 0.001) compared with Phase 1. Procedure time, PVI time, fluoroscopy time, and total applications were similar between phases. Conclusions: An impedance-change threshold of ≥ 10% on the Contact Index reliably indicates adequate contact with the pentaspline catheter and, when enforced, reduces acute PVG without prolonging procedures.

Authors

Institutions

Publication Details

Journal
Journal of Arrhythmia
Published
2026-04-01
DOI
https://doi.org/10.1002/joa3.70333
Primary Topic
Atrial Fibrillation Management and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein Gaps

Masanaru Sawada, A. Matsushima, Hikaru Masuda, Y Okumura et al.
Journal of Arrhythmia
Atrial Fibrillation Management and Outcomes
article

Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein Gaps

Masanaru Sawada, A. Matsushima, Hikaru Masuda, Y Okumura, Ryuta Watanabe, Koichi Nagashima, Shu Hirata, Yuji Saito
article en

Abstract

Background: Pulsed field ablation (PFA) for atrial fibrillation (AF) requires adequate catheter-tissue contact. The Contact Index visualizes impedance change for the pentaspline catheter, but a validated contact threshold is unknown. Objective: The objective of this study is to define an impedance-change cutoff that predicts sufficient contact to prevent acute pulmonary vein gaps (PVG) during pulmonary vein isolation (PVI), and to validate Contact-Index-guided PVI. Methods: = 26) delivered energy only when all target splines met/exceeded the cutoff value obtained from Phase 1. Results: < 0.001) compared with Phase 1. Procedure time, PVI time, fluoroscopy time, and total applications were similar between phases. Conclusions: An impedance-change threshold of ≥ 10% on the Contact Index reliably indicates adequate contact with the pentaspline catheter and, when enforced, reduces acute PVG without prolonging procedures.

Journal of ArrhythmiaVol. 42(2)
Nihon University (JP)
Openalex Percentile: Top 31%
Atrial Fibrillation Management and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.