Intraoperative Lower Extremity SSEP Absence from Over-Tightened Safety Strap: Case Report

Somatosensory evoked potentials (SSEPs) are used in spinal surgeries to monitor the functional integrity of dorsal column medial lemniscal pathways. Changes to waveform amplitudes can be due to surgical manipulation, in addition to many outside factors. We describe a posterior lumbar decompression/fusion case in which lower extremity (LE) SSEPs were performed and deteriorated bilaterally. These changes were found to be related to a safety strap traversing the posterior thighs that was excessively tight, causing neurological dysfunction. After loosening the strap, SSEP amplitudes and latencies returned to baseline parameters. This observation demonstrates the importance of careful patient positioning and application of external restraints. Numerous reports have described upper extremity (UE) positioning issues, but there are few that detail LE positioning issues. Intraoperative neuromonitoring (IONM) clinicians should be aware of various potential effects of improper positioning on IONM data and be vigilant in recognizing these instances.

Authors

Institutions

Publication Details

Journal
The Neurodiagnostic Journal
Published
2026-04-03
DOI
https://doi.org/10.1080/21646821.2026.2643816
Primary Topic
Intraoperative Neuromonitoring and Anesthetic Effects
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intraoperative Lower Extremity SSEP Absence from Over-Tightened Safety Strap: Case Report

Callan Broderick, Olivia Fisher, Emma Morley, Taylor White
The Neurodiagnostic Journal
Intraoperative Neuromonitoring and Anesthetic Effects
article

Intraoperative Lower Extremity SSEP Absence from Over-Tightened Safety Strap: Case Report

Callan Broderick, Olivia Fisher, Emma Morley, Taylor White
article en

Abstract

Somatosensory evoked potentials (SSEPs) are used in spinal surgeries to monitor the functional integrity of dorsal column medial lemniscal pathways. Changes to waveform amplitudes can be due to surgical manipulation, in addition to many outside factors. We describe a posterior lumbar decompression/fusion case in which lower extremity (LE) SSEPs were performed and deteriorated bilaterally. These changes were found to be related to a safety strap traversing the posterior thighs that was excessively tight, causing neurological dysfunction. After loosening the strap, SSEP amplitudes and latencies returned to baseline parameters. This observation demonstrates the importance of careful patient positioning and application of external restraints. Numerous reports have described upper extremity (UE) positioning issues, but there are few that detail LE positioning issues. Intraoperative neuromonitoring (IONM) clinicians should be aware of various potential effects of improper positioning on IONM data and be vigilant in recognizing these instances.

The Neurodiagnostic Journal
Seat (Spain) (ES)
Good health and well-being
Openalex Percentile: Top 4%
Intraoperative Neuromonitoring and Anesthetic Effects
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.