Abstract Introduction: Tumor Treating Fields (TTFields) are a noninvasive cancer therapy that employ alternating electric fields in the 100-500 kHz range to disrupt tumor cell division. TTFields are clinically approved for glioblastoma, mesothelioma, and non-small cell lung cancer. The LB10000 is a novel TTFields device designed to deliver high-dose fields adaptively across large body regions. Unlike existing systems that use four fixed transducer arrays, the LB10000 employs a large matrix of programmable transducers whose phases can be dynamically switched between 0°, 180°, and off. This enables dynamic field shaping, targeted delivery to multiple anatomical sites, and effective thermal management. We present both in-vivo and in-silico evidence demonstrating the safety of this new system. Methods: Six female Yucatan pigs were treated with the LB10000. Arrays were removed daily between 7-9 a.m. and reapplied between 12-3 p.m., targeting ≥16 hours of active treatment per day for 24-32 days within a period of 40 days from treatment initiation. If a pig reached the desired quota of per-protocol days, treatment was stopped and the animal euthanized. During treatment, skin temperature, treatment duty cycle, current, and voltage were continuously monitored, and animal well-being was evaluated daily by a veterinarian. Following euthanasia, gross examination of major organs was performed. To assess safety in humans, computational modeling was performed using the Sim4Life (ZMT Zurich, Switzerland) ohmic solver. Virtual LB10000 arrays were applied to anthropomorphic phantoms—DUKE (adult male), ELLA (adult female), and FATS (obese male)—and electric field delivery simulated at 200 Vpp, the device’s maximum output. Specific Absorption Rate (SAR) distributions were calculated to evaluate potential heating. Results: In vivo, the LB10000 delivered fields continuously at 130 V and 6 A (∼100 W) with 78-90 % on-time per animal. Five of six pigs completed the full protocol(24-32 day with 16hr/day treatment), totaling 2,780 hours of treatment without adverse events or tissue injury. Gross examination revealed no abnormalities. In silico, SAR values in superficial skin layers were 10-100× higher than in internal organs, indicating that any potential heating is localized beneath the transducers. The risk of thermal damage to tissues increases drastically when tissue temperatures exceed 109°F (43°C). The LB10000 continuously monitors skin temperature below the transudcers and controls delivered power to maintain skin temperature below a safety threshold of 105 °F. Together, these findings demonstrate that the LB10000 operates safely at high doses in vivo and that the risk that the device will cause thermal damage in humans is negligible. Conclusion: These results provide key evidence supporting advancement toward first-in-human clinical evaluation. Citation Format: Ze'ev Bomzon, Scott Krywick, Matthew Travers, Kenneth L. Watkins, Martin Pribula, Michael Winegar, Peter Travers. From simulation to reality: Evidence supporting safe high-dose TTFields delivery with the LB10000 [abstract]. In: Proceedings of the American Association for Cancer Research Annual Meeting 2026; Part 1 (Regular Abstracts); 2026 Apr 17-22; San Diego, CA. Philadelphia (PA): AACR; Cancer Res 2026;86(7 Suppl):Abstract nr 7199.

Authors

Institutions

Publication Details

Journal
Cancer Research
Published
2026-04-03
DOI
https://doi.org/10.1158/1538-7445.am2026-7199
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 7199: From simulation to reality: Evidence supporting safe high-dose TTFields delivery with the LB10000

Kenneth Watkins, Martin Pribula, Ze'ev Bomzon, Scott Krywick et al.
Cancer Research
Ultrasound and Hyperthermia Applications
article

Abstract 7199: From simulation to reality: Evidence supporting safe high-dose TTFields delivery with the LB10000

Kenneth Watkins, Martin Pribula, Ze'ev Bomzon, Scott Krywick, Matthew Travers, Michael Winegar, Peter Travers
article en

Abstract

Abstract Introduction: Tumor Treating Fields (TTFields) are a noninvasive cancer therapy that employ alternating electric fields in the 100-500 kHz range to disrupt tumor cell division. TTFields are clinically approved for glioblastoma, mesothelioma, and non-small cell lung cancer. The LB10000 is a novel TTFields device designed to deliver high-dose fields adaptively across large body regions. Unlike existing systems that use four fixed transducer arrays, the LB10000 employs a large matrix of programmable transducers whose phases can be dynamically switched between 0°, 180°, and off. This enables dynamic field shaping, targeted delivery to multiple anatomical sites, and effective thermal management. We present both in-vivo and in-silico evidence demonstrating the safety of this new system. Methods: Six female Yucatan pigs were treated with the LB10000. Arrays were removed daily between 7-9 a.m. and reapplied between 12-3 p.m., targeting ≥16 hours of active treatment per day for 24-32 days within a period of 40 days from treatment initiation. If a pig reached the desired quota of per-protocol days, treatment was stopped and the animal euthanized. During treatment, skin temperature, treatment duty cycle, current, and voltage were continuously monitored, and animal well-being was evaluated daily by a veterinarian. Following euthanasia, gross examination of major organs was performed. To assess safety in humans, computational modeling was performed using the Sim4Life (ZMT Zurich, Switzerland) ohmic solver. Virtual LB10000 arrays were applied to anthropomorphic phantoms—DUKE (adult male), ELLA (adult female), and FATS (obese male)—and electric field delivery simulated at 200 Vpp, the device’s maximum output. Specific Absorption Rate (SAR) distributions were calculated to evaluate potential heating. Results: In vivo, the LB10000 delivered fields continuously at 130 V and 6 A (∼100 W) with 78-90 % on-time per animal. Five of six pigs completed the full protocol(24-32 day with 16hr/day treatment), totaling 2,780 hours of treatment without adverse events or tissue injury. Gross examination revealed no abnormalities. In silico, SAR values in superficial skin layers were 10-100× higher than in internal organs, indicating that any potential heating is localized beneath the transducers. The risk of thermal damage to tissues increases drastically when tissue temperatures exceed 109°F (43°C). The LB10000 continuously monitors skin temperature below the transudcers and controls delivered power to maintain skin temperature below a safety threshold of 105 °F. Together, these findings demonstrate that the LB10000 operates safely at high doses in vivo and that the risk that the device will cause thermal damage in humans is negligible. Conclusion: These results provide key evidence supporting advancement toward first-in-human clinical evaluation. Citation Format: Ze'ev Bomzon, Scott Krywick, Matthew Travers, Kenneth L. Watkins, Martin Pribula, Michael Winegar, Peter Travers. From simulation to reality: Evidence supporting safe high-dose TTFields delivery with the LB10000 [abstract]. In: Proceedings of the American Association for Cancer Research Annual Meeting 2026; Part 1 (Regular Abstracts); 2026 Apr 17-22; San Diego, CA. Philadelphia (PA): AACR; Cancer Res 2026;86(7 Suppl):Abstract nr 7199.

Cancer ResearchVol. 86(7_Supplement)
University of Minnesota Health Maple Grove Clinics (US), Luna Innovations (United States) (US)
Good health and well-being
Openalex Percentile: Top 32%
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.