Validation of a Low-Volume (100 μL) Plasma Protocol for HIV-1 RNA Quantification Using the Hologic Aptima HIV-1 Quant Dx Assay
copies/mL. No false-positive or false-negative results were observed among 30 fully suppressed samples. Precision analyses showed minimal variability, with all results within predefined acceptability criteria. These findings indicate that the low-volume Aptima protocol provides accurate and reliable HIV-1 RNA quantification and can expand access to viral load testing in pediatric patients and other clinical settings in which plasma volume is limited.
Authors
- Lorna Madurai
- Kayla Pillay
- Bhavani Manivannan
- Kiveshan Reddy
- Someshni Nair
- Sharana Mahomed
Institutions
- Centre for the AIDS Programme of Research in South Africa (ZA)
- Universal University (RU)
- VRR Diagnostic Services (India) (IN)
Publication Details
- Journal
- Journal of Molecular Diagnostics
- Published
- 2026-04-01
- DOI
- https://doi.org/10.1016/j.jmoldx.2026.03.006
- Primary Topic
- HIV Research and Treatment
- Type
- article
- Field-Weighted Citation Impact
- 0.00