Using a physics-based approach to standardize radiomics values: experimental validation in an anthropomorphic phantom on a clinical CT scanner using a range of dose levels and reconstruction kernels

Radiomics relies on quantitative features to discern the underlying biological signatures. However, feature dependence on the imaging systems themselves hampers the creation of reproducible and generalizable models. We have previously proposed a novel framework to remove the effects of system blur and image noise on radiomic calculations and performed validation in simulation studies. In this work, we extended the analysis and evaluated the method on CT data acquired of an anthropomorphic phantom with realistic lung textures. Data was acquired at five different dose levels and reconstructed using eight different reconstruction kernels. To test the generalizability of the method, we applied our proposed method to standardize from all possible starting and reference kernel pairs under all measured dose levels for a total of 320 cases (8×8×5). Standardization was performed for radiomics features from four classes, histogram, GLCM, gray-level run length matrix-(GLRLM), and wavelet transforms. Results indicate that standardized radiomics features are closer to the reference and on average, the average absolute percentage difference from reference over all features is improved by a factor of three compared with unstandardized features. In addition, we found that standardization from a smoother to a sharper kernel is a more challenging task and that performance is comparable across all dose levels. This work shows that the proposed standardization method is effective in standardizing radiomics feature values across a wide range of imaging conditions in clinical CT.

Authors

Institutions

Publication Details

Published
2026-04-02
DOI
https://doi.org/10.1117/12.3087679
Primary Topic
Radiomics and Machine Learning in Medical Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Using a physics-based approach to standardize radiomics values: experimental validation in an anthropomorphic phantom on a clinical CT scanner using a range of dose levels and reconstruction kernels

Huay Din, Yijie Yuan, GRACE HYUN KIM, Michael F. McNitt-Gray et al.
Radiomics and Machine Learning in Medical Imaging
article

Using a physics-based approach to standardize radiomics values: experimental validation in an anthropomorphic phantom on a clinical CT scanner using a range of dose levels and reconstruction kernels

Huay Din, Yijie Yuan, GRACE HYUN KIM, Michael F. McNitt-Gray, Grace J. Gang, J. Webster Stayman
article en

Abstract

Radiomics relies on quantitative features to discern the underlying biological signatures. However, feature dependence on the imaging systems themselves hampers the creation of reproducible and generalizable models. We have previously proposed a novel framework to remove the effects of system blur and image noise on radiomic calculations and performed validation in simulation studies. In this work, we extended the analysis and evaluated the method on CT data acquired of an anthropomorphic phantom with realistic lung textures. Data was acquired at five different dose levels and reconstructed using eight different reconstruction kernels. To test the generalizability of the method, we applied our proposed method to standardize from all possible starting and reference kernel pairs under all measured dose levels for a total of 320 cases (8×8×5). Standardization was performed for radiomics features from four classes, histogram, GLCM, gray-level run length matrix-(GLRLM), and wavelet transforms. Results indicate that standardized radiomics features are closer to the reference and on average, the average absolute percentage difference from reference over all features is improved by a factor of three compared with unstandardized features. In addition, we found that standardization from a smoother to a sharper kernel is a more challenging task and that performance is comparable across all dose levels. This work shows that the proposed standardization method is effective in standardizing radiomics feature values across a wide range of imaging conditions in clinical CT.

Vol. 13924
Johns Hopkins University (US), University of California, Los Angeles (US), University of Pennsylvania (US)
National Institutes of Health
Openalex Percentile: Top 22%
Radiomics and Machine Learning in Medical Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.