Experimental Determination of the 12C(12C, X)11C Reaction Cross Sections for Intensity Monitor Calibration of an Extracted Carbon Ion Beam of Clinical Energies

At present, National Research Center “Kurchatov Institute”—IHEP has developed an original technique for absolute calibration of ionization chambers with activation detectors, when working with extracted beams of accelerated carbon ions of clinical energies of 150–450 MeV/nucleon. A low-background radiometric setup operating on the principle of γ–γ coincidences and an activation detector made of pure carbon with natural mixture of isotopes and the 12C(12C, X)11C reaction are used. To increase the accuracy of this technique, experimental values of the cross section of the 12C(12C, X)11C reaction at energies of 450 and 400 MeV/nucleon have been obtained at the U-70 accelerator complex. A comparison with existing world data is performed, and it is shown that the result obtained is in satisfactory agreement with them.

Authors

Institutions

Publication Details

Journal
Physics of Particles and Nuclei
Published
2026-04-01
DOI
https://doi.org/10.1134/s1063779625701205
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Experimental Determination of the 12C(12C, X)11C Reaction Cross Sections for Intensity Monitor Calibration of an Extracted Carbon Ion Beam of Clinical Energies

M. Yu. Kostin, A. A. Yanovich, A. G. Vasilyeva, O. V. Sumaneev et al.
Physics of Particles and Nuclei
Radiation Therapy and Dosimetry
article

Experimental Determination of the 12C(12C, X)11C Reaction Cross Sections for Intensity Monitor Calibration of an Extracted Carbon Ion Beam of Clinical Energies

M. Yu. Kostin, A. A. Yanovich, A. G. Vasilyeva, O. V. Sumaneev, V. A. Pikalov
article en

Abstract

At present, National Research Center “Kurchatov Institute”—IHEP has developed an original technique for absolute calibration of ionization chambers with activation detectors, when working with extracted beams of accelerated carbon ions of clinical energies of 150–450 MeV/nucleon. A low-background radiometric setup operating on the principle of γ–γ coincidences and an activation detector made of pure carbon with natural mixture of isotopes and the 12C(12C, X)11C reaction are used. To increase the accuracy of this technique, experimental values of the cross section of the 12C(12C, X)11C reaction at energies of 450 and 400 MeV/nucleon have been obtained at the U-70 accelerator complex. A comparison with existing world data is performed, and it is shown that the result obtained is in satisfactory agreement with them.

Physics of Particles and NucleiVol. 57(2)
Kurchatov Institute (RU), Institute for High Energy Physics (RU)
Openalex Percentile: Top 27%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Experimental Determination of the 12C(12C, X)11C Reaction Cross Sections for Intensity Monitor Calibration of an Extracted Carbon Ion Beam of Clinical Energies — M. Yu. Kostin, A. A. Yanovich, et al. · Physics of Particles and Nuclei (2026) | TGRS Research Map | TGRS