Safety Testing of Endovascular Devices In Vitro for Interventional Neuroradiology Under 0.55 T MRI
Background/Objectives: MRI-guided neurovascular interventions could benefit from lower-field systems due to reduced magnetic and radiofrequency hazards. However, safety and practical visibility of commonly used neurointerventional devices at 0.55 T remain insufficiently characterized. We evaluated magnetic field interactions, RF-induced heating, and qualitative device visibility in 11 commercially available and commonly used neurovascular devices on a 0.55 T MRI system. Methods: Eleven devices, including stent retrievers, guidewires, catheters, and one embolization implant, were tested at 0.55 T. Magnetostatic interactions were quantified using the American Society for Testing and Materials (ASTM)-guided deflection methods for translational force (ASTM-F2052) and a two-string suspension apparatus for torque (adapted from Stoianovici et al.). RF-induced heating was measured in an in vitro perfused cerebral vessel phantom using a 15 min high-specific absorption rate spin echo sequence under static and flow conditions. Qualitative device visibility was assessed using a turbo spin echo (TSE) and balanced steady-state free precession (bSSFP) imaging on each device individually. Results: Eight of eleven devices passed the translational force test, while three devices (D, E, and G), containing significant ferromagnetic components, failed with deflection angles > 45°. Eight devices passed torque testing, remaining below the critical threshold in all rotation positions; three devices (D, G, and J) failed by exceeding the 54° criterion, including one guidewire and two devices with braided/coiled metallic structures. Under static conditions, RF-induced heating ranged from negligible to 10.4 °C (maximum in device D) and generally decreased under flow; in the flow configuration, temperature rise remained below 2 °C for 6/11 devices. Qualitative imaging performance differed by sequence, with bSSFP enabling improved delineation of device structure (best for devices A, C, and H), whereas devices D, E, F, and J produced extensive signal voids that precluded reliable visualization in both sequences. Overall, three devices satisfied all safety criteria while remaining clearly visible under MRI. Conclusions: Devices that pass safety thresholds at 0.55 T can serve as candidates for further sequence optimization and preclinical workflow development, enabling the design of low-SAR, device-compatible imaging protocols tailored for neurointerventional workflows. These results provide key safety data supporting the feasibility of MR-guided neurovascular procedures at 0.55 T.
Authors
- Steven Hajdu (ORCID: https://orcid.org/0000-0001-7659-4325)
- Guillaume Saliou (ORCID: https://orcid.org/0000-0003-3832-7976)
- Katerina Eyre (ORCID: https://orcid.org/0000-0003-4208-6140)
- Bruno Bartolini (ORCID: https://orcid.org/0000-0001-6055-0221)
- Stanislas Rapacchi (ORCID: https://orcid.org/0000-0002-8925-495X)
- Francesco Puccinelli (ORCID: https://orcid.org/0000-0002-9171-8063)
- Karolina Swierdzewska
- Jean‐Baptiste Ledoux (ORCID: https://orcid.org/0000-0003-0447-5073)
- Adèle L.C. Mackowiak
- Matthias Stuber
- Christopher W. Roy
Institutions
- Centre d'Imagerie BioMedicale (CH)
- University of Lausanne (CH)
Publication Details
- Journal
- Neuroimaging
- Published
- 2026-04-02
- DOI
- https://doi.org/10.3390/neuroimaging1020007
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00