The impact of surface dimensions on the accuracy of an intraoral scanner

OBJECTIVES: Intra-oral scanners are being used to measure small changes to the surface form. This study aimed to identify the limits of detection of an intra-oral scanner (IOS) on depth and width of surface form. METHODS: We performed investigations on artificial teeth with increasing width and depth of discrete surface features captured by an intra-oral scanner at baseline and compared the data to a gold-standard white-light profilometer (WLP). 81 surface features with widths varying between 150 - 1300 µm and depths 5 - 360 µm were scanned five times using the WLP (n = 405) and IOS (n = 405) and analysed using a validated registration and subtraction algorithm. Both scan outputs used reference and an iterative closest point (ICP) registration to the Z-axis to analyse change and control variation of data. 2D maximum step height depth (µm) and 2D width (µm) of the surface features were calculated. RESULTS: = 0.76). The error of depth measurement was 100 % at 250 µm, 20 % at 750 µm and 0 % at 1200 µm. Overall, the IOS overestimated surface feature width by 155.5 µm and underestimated surface feature depth by -20.2 µm. CONCLUSION: The accuracy of the IOS to capture surface features improved above 750 µm in width and was close to 100 % accuracy at 1200 µm. Caution is needed if an IOS is used to measure surface features below 1200 µm in width.

Authors

Institutions

Publication Details

Journal
Journal of Dentistry
Published
2026-04-02
DOI
https://doi.org/10.1016/j.jdent.2026.106669
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The impact of surface dimensions on the accuracy of an intraoral scanner

Thomas Bull, Rupert Austin, Saoirse O'Toole, Sneha Chotaliya et al.
Journal of Dentistry
Dental Radiography and Imaging
article

The impact of surface dimensions on the accuracy of an intraoral scanner

Thomas Bull, Rupert Austin, Saoirse O'Toole, Sneha Chotaliya, David Bartlett
article en

Abstract

OBJECTIVES: Intra-oral scanners are being used to measure small changes to the surface form. This study aimed to identify the limits of detection of an intra-oral scanner (IOS) on depth and width of surface form. METHODS: We performed investigations on artificial teeth with increasing width and depth of discrete surface features captured by an intra-oral scanner at baseline and compared the data to a gold-standard white-light profilometer (WLP). 81 surface features with widths varying between 150 - 1300 µm and depths 5 - 360 µm were scanned five times using the WLP (n = 405) and IOS (n = 405) and analysed using a validated registration and subtraction algorithm. Both scan outputs used reference and an iterative closest point (ICP) registration to the Z-axis to analyse change and control variation of data. 2D maximum step height depth (µm) and 2D width (µm) of the surface features were calculated. RESULTS: = 0.76). The error of depth measurement was 100 % at 250 µm, 20 % at 750 µm and 0 % at 1200 µm. Overall, the IOS overestimated surface feature width by 155.5 µm and underestimated surface feature depth by -20.2 µm. CONCLUSION: The accuracy of the IOS to capture surface features improved above 750 µm in width and was close to 100 % accuracy at 1200 µm. Caution is needed if an IOS is used to measure surface features below 1200 µm in width.

Journal of DentistryVol. 170
Guy's Hospital (GB), King's College London (GB), King's College Hospital (GB), University of Southampton (GB)
Medical Research Council
Openalex Percentile: Top 35%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.