Fluence-corrected pencil beam algorithm for carbon ion therapy patient quality assurance in solid phantoms

Solid phantoms streamline carbon ion therapy quality assurance (QA), but introduce dose deviations as treatment planning system's (TPS) pencil beam algorithms (PBA) calculate dose in water. This study develops a fluence-corrected PBA, enabling accurate patient-specific QA directly in solid phantoms at Heavy Ion Medical Machine (HIMM) centers. Beam modeling was done in TOPAS Monte Carlo simluation platform to match the integrated depth doses (IDD) and phase space parameters to the clinical commissioning data. Depth-dose data in PTW RW3 slab phantom and in water were simulated, from which energy- and depth-dependent fluence correction (FC) factors (FCFs) were derived. The FCFs were validated experimentally and incorporated into the PBA. 4 types of QA workflows were compared for 2 patient plans at varying equivalent depths: 1) phantom CT without FC-PBA 2) phantom CT with FC 3) virtual CT with stopping power ratio overwritten to 1 with FC, and 4) virtual CT overwritten with phantom SPR with FC. A simplified depth-dependent correction table was fitted from measurement. The simulated FCFs increased with energy and depth from about 0.4% at 2 centimeters to 5% at 25 centimeters, which agreed with the measurement within ±0.5%. At depths < 100 mm, the gamma pass rates were similar for all 4 workflows. For depths > 100 mm, the gamma pass rates increased from (81, 77, 69)% for workflow 1 to (99, 99, 97)% on average for workflows 2-4, under 3 mm/3%, 2 mm/3% and 2 mm/2% criteria, respectively. Workflow 2 and 4 gave similar results with mean difference 0.1(±0.6)%. The simplified table gave 0.0231% FC per millimeter of RW3 thickness, which agreed to the simulation within 1%. Carbon ion therapy QA in the RW3 requires FCFs particularly for > 100 mm depths. Under ≥ 2% pass rate criteria, either the fluence-corrected PBA or the simplified table could be adopted on a virtual phantom to support easier QA workflow.

Authors

Institutions

Publication Details

Journal
Medical dosimetry
Published
2026-04-01
DOI
https://doi.org/10.1016/j.meddos.2026.02.003
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fluence-corrected pencil beam algorithm for carbon ion therapy patient quality assurance in solid phantoms

Hongfei Li, Yunzhou Xia, Xiuzhen Deng, Qinqin Cheng et al.
Medical dosimetry
Radiation Therapy and Dosimetry
article

Fluence-corrected pencil beam algorithm for carbon ion therapy patient quality assurance in solid phantoms

Hongfei Li, Yunzhou Xia, Xiuzhen Deng, Qinqin Cheng, Kai Wang
article en

Abstract

Solid phantoms streamline carbon ion therapy quality assurance (QA), but introduce dose deviations as treatment planning system's (TPS) pencil beam algorithms (PBA) calculate dose in water. This study develops a fluence-corrected PBA, enabling accurate patient-specific QA directly in solid phantoms at Heavy Ion Medical Machine (HIMM) centers. Beam modeling was done in TOPAS Monte Carlo simluation platform to match the integrated depth doses (IDD) and phase space parameters to the clinical commissioning data. Depth-dose data in PTW RW3 slab phantom and in water were simulated, from which energy- and depth-dependent fluence correction (FC) factors (FCFs) were derived. The FCFs were validated experimentally and incorporated into the PBA. 4 types of QA workflows were compared for 2 patient plans at varying equivalent depths: 1) phantom CT without FC-PBA 2) phantom CT with FC 3) virtual CT with stopping power ratio overwritten to 1 with FC, and 4) virtual CT overwritten with phantom SPR with FC. A simplified depth-dependent correction table was fitted from measurement. The simulated FCFs increased with energy and depth from about 0.4% at 2 centimeters to 5% at 25 centimeters, which agreed with the measurement within ±0.5%. At depths < 100 mm, the gamma pass rates were similar for all 4 workflows. For depths > 100 mm, the gamma pass rates increased from (81, 77, 69)% for workflow 1 to (99, 99, 97)% on average for workflows 2-4, under 3 mm/3%, 2 mm/3% and 2 mm/2% criteria, respectively. Workflow 2 and 4 gave similar results with mean difference 0.1(±0.6)%. The simplified table gave 0.0231% FC per millimeter of RW3 thickness, which agreed to the simulation within 1%. Carbon ion therapy QA in the RW3 requires FCFs particularly for > 100 mm depths. Under ≥ 2% pass rate criteria, either the fluence-corrected PBA or the simplified table could be adopted on a virtual phantom to support easier QA workflow.

Medical dosimetry
Zhejiang Academy of Social Sciences (CN), Lanzhou City University (CN), China Academy of Chinese Medical Sciences (CN)
Openalex Percentile: Top 22%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.