A scaling law for hemolysis and rotating speed in rotary blood pumps

This study presents the first demonstration that a scaling law for red blood cell damage (hemolysis) modeling in rotary blood pumps exists that allows for the prediction of blood damage under any operating condition from the characteristic hemolysis curve at an arbitrary pump speed. This new scaling law was validated for normalized measures of hemolysis, including Modified Index of Hemolysis (MIH), for both computational simulations and experimental measurements, in multiple different rotary blood pumps. The ratio of MIH was related to the ratio of rotating speed, ω, such that (MIH1/MIH2)=(ω1/ω2)n. The coefficient, n, depends on pump geometry, and it has been shown that n can be determined with a single additional test or simulation. Previously, the prediction of hemolysis at any desired pump operating point required experimental testing or computational simulation at that specific condition, resulting in a lengthy and expensive design process. The functionality of the scaling law was demonstrated in the design of a novel proposed pediatric blood pump. The blade shapes were optimized for minimal hemolysis both with and without the use of the scaling law. The methods resulted in identical pumps for well-constrained optimization with computational expense halved using hemolysis scaling. Moreover, no additional simulations are required for further off-design blood damage analysis using this new method. With healthcare product development costs rising unsustainably, this new hemolysis scaling law has the potential to drastically reduce the cost of innovation and accelerate the time to market for rotary blood pumps, maximizing the life-saving potential of this technology.

Authors

Institutions

Publication Details

Journal
Physics of Fluids
Published
2026-04-01
DOI
https://doi.org/10.1063/5.0301568
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A scaling law for hemolysis and rotating speed in rotary blood pumps

Lee Nissim, N. S. Kelly, K. H. Fraser, O. H. Frazier et al.
Physics of Fluids
Mechanical Circulatory Support Devices
article

A scaling law for hemolysis and rotating speed in rotary blood pumps

Lee Nissim, N. S. Kelly, K. H. Fraser, O. H. Frazier, C. H. H. Chan, Y. Wang
article en

Abstract

This study presents the first demonstration that a scaling law for red blood cell damage (hemolysis) modeling in rotary blood pumps exists that allows for the prediction of blood damage under any operating condition from the characteristic hemolysis curve at an arbitrary pump speed. This new scaling law was validated for normalized measures of hemolysis, including Modified Index of Hemolysis (MIH), for both computational simulations and experimental measurements, in multiple different rotary blood pumps. The ratio of MIH was related to the ratio of rotating speed, ω, such that (MIH1/MIH2)=(ω1/ω2)n. The coefficient, n, depends on pump geometry, and it has been shown that n can be determined with a single additional test or simulation. Previously, the prediction of hemolysis at any desired pump operating point required experimental testing or computational simulation at that specific condition, resulting in a lengthy and expensive design process. The functionality of the scaling law was demonstrated in the design of a novel proposed pediatric blood pump. The blade shapes were optimized for minimal hemolysis both with and without the use of the scaling law. The methods resulted in identical pumps for well-constrained optimization with computational expense halved using hemolysis scaling. Moreover, no additional simulations are required for further off-design blood damage analysis using this new method. With healthcare product development costs rising unsustainably, this new hemolysis scaling law has the potential to drastically reduce the cost of innovation and accelerate the time to market for rotary blood pumps, maximizing the life-saving potential of this technology.

Physics of FluidsVol. 38(4)
Texas Medical Center (US), Institute for Biotechnology and Bioengineering (PT), University of Bath (GB)
National Heart, Lung, and Blood Institute
Industry, innovation and infrastructure
Openalex Percentile: Top 26%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A scaling law for hemolysis and rotating speed in rotary blood pumps — Lee Nissim, N. S. Kelly, et al. · Physics of Fluids (2026) | TGRS Research Map | TGRS